UF-1000i尿沉渣分析仪在排除尿路感染中的应用评价 被引量:4

Evaluation of the Application of UF-1000i Automatic Urine Dreg Analyzer in Excluding Urinary Tract Infection
原文传递
导出
摘要 目的评价UF-1000i全自动尿沉渣分析仪中沉渣定量模块中白细胞和细菌参数阈值在排除尿路感染的应用价值。方法选取2 580份清洁中段尿液,同时进行细菌培养菌落计数和UF-1000i尿沉渣白细胞和细菌定量分析,建立ROC曲线确定白细胞参数与细菌参数阈值。结果以尿定量培养菌落计数G菌≥105 cfu/mL,G+菌≥104 cfu/mL为阳性参考标准,当白细胞沉渣定量为100/μL时,UF-1000i尿沉渣分析仪检测灵敏度为64%,特异度为75%,阴性预测值为96%;当UF-1000i细菌计量为901/μL时,检测灵敏度为68.3%,特异度为92.8%,阴性预测值为97%。结论 UF-1000i检测新鲜尿标本白细胞的测定值<100/μL,细菌值<901/μL时能够作为临床早期排除尿路感染的依据之一。 Objective To evaluate the application value of the quantitative module of UF-1000i automatic urine dreg analyzer in excluding urinary infection by the thresholds of white blood cells(WBCs) and bacterial parameters.Methods A total of 2 580 clean urine samples were collected to simultaneously perform urine quantitative cultures and quantitative analysis of the white blood cells and bacteria by UF-1000i urine dreg,and then ROC curves were established to determine the threshold of white blood cells and bacteria parameters.Results According to the reference index that no less than 105 colonies for gram-negative bacterium and 104 colonies for gram-positive bacterium were detected by urine quantitative cultures,while cut-off value of WBCs was 100/礚,a sensitivity of 64%,a specificity of 75%,and a negative predictive value(NPV) of 96% were obtained;while cut-off value of bacteria was 901/礚,a sensitivity of 68.3%,a specificity of 92.8 %,and an NPV of 97% were obtained.Conclusion The cut-off values of 901/礚 for bacteria and 100/礚 for WBCs in the urine specimen determined by UF-1000I can be one of the bases for early exclusion of urinary tract infections.
出处 《华西医学》 CAS 2013年第7期1033-1035,共3页 West China Medical Journal
关键词 尿路感染 UF-1000i尿沉渣分析仪 阈值 Urinary tract infection UF-1000I urine dreg analyzer Cut-off value
  • 相关文献

参考文献11

  • 1黎国庆,郭明好,杨磊,刘云.膀胱穿刺尿同步细菌培养在女性尿路感染诊断中的价值[J].郑州大学学报(医学版),2002,37(2):224-225. 被引量:5
  • 2Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures[J]. J Clin Microbiol, 2001, 39(11): 4169-4171.
  • 3Kouri TT, K:lk6nen U, Malminiemi K, et al. Evaluation of sysmex UF-100 urine flow cytometer vs chamber counting of supravitally stained specimens and conventional bacterial cultures[J]. Am J Clin Pathol, 1999, 112(1): 25-35.
  • 4魏梦岚,陈伟,高洁.UF-100尿沉渣分析仪在尿路感染早期诊断中的应用[J].现代医药卫生,2008,24(4):483-484. 被引量:3
  • 5宁永忠,马纪志,张捷.全自动尿沉渣分析仪UF-100中细菌计数参数阈值的研究[J].中国误诊学杂志,2009,9(36):8824-8826. 被引量:1
  • 6Koken T, Aktepe OC, Serteser M, et al. Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF-100) in urinary tract infections[J]. Int Urol Nephrol, 2002, 34(2): 175-178.
  • 7Evans R, Davidson MM, Sire LR, et al. Testing by sysmex UF-100 flow cytometer and with bacterial culture in a diagnostic laboratory: a comparison[J]. J Clin Pathol, 2006, 59(6): 661-662.
  • 8Fenili D, Pirovano B. The automation of sediment urinalysis using a new urine flow cytometer (UF-100)[J]. Clin Chem Lab Med, 1998, 36(12): 909-917.
  • 9马筱玲,李桂琴.临床微生物学尿培养操作规范[J].中华检验医学杂志,2005,28(10):1085-1087. 被引量:38
  • 10Rosa RD, Grosso S, Bruschetta G, et al. Evaluation of the sysmex UF1000i flow cytometer for ruling out bacterial urinary tract infection[J]. Clin Chim Acta, 2010, 411(15-16): 1137-1142.

二级参考文献26

  • 1金大鸣.UF-100型尿沉渣全自动检测仪评价[J].中华检验医学杂志,1998,22(3):61-62. 被引量:145
  • 2刘华,王颖,钟亚玲,李焱鑫,喻华.UF-100尿沉渣分析仪细菌计数与尿细菌培养的比较[J].现代预防医学,2005,32(5):497-498. 被引量:8
  • 3李广华,陈雄雁,罗燕飞,叶联珍.联合UF-100型尿沉渣分析仪和尿干化学分析诊断尿路感染的研究[J].实用医学杂志,2007,23(7):1068-1069. 被引量:6
  • 4叶任高 张仕光.尿路感染诊断的研究[J].中华内科杂志,1986,25:400-400.
  • 5U. S. Preventive Services Task Force. Screening for Asymptomatic Bacteriuria in Adults:U. S. Preventive Services Task Force Reaffirmation Recommendation Statement [J]. Ann Intern Med, 2008,149:43-47.
  • 6Bent S, Nallamothu BK,Simel DL,et al. Does this woman have an acute uncomplicated urinary tract infection[J]. JAMA, 2002,287 (20) :2701-2710.
  • 7Mayo S,Acevedo D,Quinones-Torrelo C,et al. Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments[J]. J Clin Lab Anal, 2008,22 (4):262-270.
  • 8Ito K,Tatsumi N, Kikuchi H, et al. Basic evaluation of the new fully automated urine sediment analyzer (U-FCM) based on flowcytometric technology[J]. Rinsho Byori, 1994,42 (10): 1093- 1098.
  • 9KimSY,KimYJ,Lee SM,et al. Evaluation of the SysmexUF 100 urine cell analyzer as a screening test to reduce the need for urine cultures for community-acquired urinary tract infection[J].Am J Clin Pathol,2007,128(6):922 925.
  • 10dos Santos JC,Weber LP,Perez LR. Evaluation of urinalysis parameters to predict urinary-tract infection[J]. Braz J Infect Dis, 2007,11 (5) :479-481.

共引文献47

同被引文献33

引证文献4

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部