期刊文献+

基于C^4D技术的小通道流体流量测量装置及方法 被引量:1

A Device and Method For Flowrate Measurement of Fluid in Small Channel Based on C^4D Technique
原文传递
导出
摘要 目前适用于小通道的流量测量装置或仪表还十分有限。本文基于电容耦合式非接触电导测量(Capacitively Coupled Contactless Conductivity Detection,C^4D)技术和互相关测速原理,提出了一种毫米级管道的流量测量新方法,并研制了其测量装置.该方法利用两对电子开关对两路测量通路交互闭合,分别获得两组流体电导信号。对得到的两组信号进行互相关运算,从而获得流体的流速和流量.作为初步研究,本文分别对内径为3.2 mm和5.4 mm的管道进行了流量测量实验。实验结果表明基于C^4D技术的小通道流体流量测量新方法是可行的和有效的。 The flowrate measurement devices or instruments in small channel are very limited. Based on Capacitively Coupled Contactless Conductivity Detection (C4D) technique and cross-correlation velocity measurement principle, a new method for flowrate measurement in millimeter-scale pipe is proposed and a new measurement device is developed. Two pairs of electronic switches are used to obtain two independent conductivity signals of the fluid. By cross-correlating the two signals, the velocity and flowrate of the fluid can be determined. As a preliminary research, flowrate measurement experiments are carried out in two pipes with the inner diameters of 3.2 mm and 5.4 mm, respectively. The experimental results show that the new proposed method based on C4D technique for flowrate measurement is feasible and effective.
机构地区 浙江大学
出处 《工程热物理学报》 EI CAS CSCD 北大核心 2013年第9期1679-1682,共4页 Journal of Engineering Thermophysics
基金 国家自然科学基金资助项目(No.51076141 No.61074173)
关键词 电容耦合式非接触电导测量 串联谐振 电子开关 流量测量 Capacitively Coupled Contactless Conductivity Detection series resonance electronic switch flowrate measurement
  • 相关文献

参考文献8

  • 1Kandlikar SG.Fundamental Issues Related to Flow Boiling in Minichannels and Microchannels Exp [J].Therm Fluid Sci,2002,26:38-47.
  • 2F Laugere.On-Chip Contactless Four-Electrode Conductivity Detection for Capillary Electrophoresis Devices [J].Anal Chem 2003,75:306-312.
  • 3F Laugere.Design of an Electronic Interface for Capaci-tively Coupled Four-Electrode Conductivity Detection in Capillary Electrophoresis Microchip [J].Sensors and Actuators B,2002,83:104-108.
  • 4A J Zemann,E Schnell,D Volgger,G K Bonn.Contactless Conductivity Detection for Capillary Electrophoresis [J],Anal Chem,1998,70:563-567.
  • 5JAF da Silva,C L do Lago.An Oscillometric Detector for Capillary Electrophoresis [J].Anal Chem,1998,70:4339-4343.
  • 6P Kuban,P C Hauser.A Review of the Recent Achievements in Capacitively Coupled Contactless Conductivity Detection [J].Anal Chim Acta,2008,607:15-29.
  • 7HUANG Zhiyao,JIANG Weiwei,ZHOU Xinmiao,WANG Baoliang,JI Haifeng,LI Haiqing.A New Method of Capacitively Coupled Contactless Conductivity Detection Based on Series Resonance [J].Sensors and Actuators B:Chemical,2009,143(1):239-245.
  • 8HUANG Zhiyao,LONG Jun,XU Wenbo,et al.Design of Capacitively Coupled Contactless Conductivity Detection Sensor [J].Flow Measurement and Instrumentation,2012,27:67-70.