期刊文献+

具有柔韧性衬底的视网膜微电极阵列的设计 被引量:1

Design and Fabrication of Flexible Epi-retinal Microelectrode
下载PDF
导出
摘要 目的:针对视觉假体关键部件视网膜微电极阵列的结构特点和植入要求,设计并制作表层视网膜微电极阵列,并对微电极阵列特性进行测试。方法:以parylene作为柔性衬底材料,Pt为电极阵列材料,设计并制作单层电极及引线分布的视网膜植入微电极阵列,并用三电极测试系统对制作完成的微电极阵列进行电特性测试。结果:对微电极阵列的形态和电化学特性测试结果表明,制作的微电极阵列在外观形态和电学特性上符合设计和视网膜植入的要求。结论:该微电极阵列可以为进一步的动物体电生理实验提供性能可靠的视网膜微电极阵列。 Objectives To design and fabricate implantable epi-retinal microelectrode array for the purpose of retinal repair, and perform test on the features of the array, biethotls With parylene as flexible substrate material and Pt as electrode material, monolayer microelectrode array was designed and fabricated, and electric characteristics of the array were measured with the three-electrode test system. Results Test results showed that morphological and electrochemical properties of the array met the design requirements. Conclusion The microelectrode array can be put in the in vivo electrophysiological experiments on anima.
出处 《医疗卫生装备》 CAS 2013年第8期1-3,共3页 Chinese Medical Equipment Journal
基金 天津市科技支撑计划重点项目(10zckfsy08400) 国家自然科学基金项目(81271667)
关键词 视觉假体 微电极阵列 电学特性 visual prosthesis microelectrode array electrical characteristics
  • 相关文献

参考文献14

  • 1Weiland J D,Cho A K, Humayun M S. Retinal prostheses :current clinical results and future needs[J]. Ophthalmology, 2011,118 ( 11 ) : 2 227-2 237.
  • 2Klanke S,Goertz M,Rein S,et a/. Stimulation with a wireless intraocu- lar epiretinal implant elicits visual percepts in blind humans[J]. Invest Ophthalmol Vis Sci, 2011,52 ( 1 ) : 449-455.
  • 3RAO C,YUAN X H,ZHANC S J,et al. Epi-retinal prosthesisi for outer retinal degenerative diseases[J]. Intenational Journal of Oph- thamology, 2005,5 (5) : 847-851.
  • 4Chader G J, Weiland J, Humayun M S. Artificial vision:needs, func- tioning,and testing of a retinal electronic prosthesis[J]. Progress in Brain Research, 2009,175 : 317-332.
  • 5Stieglitz T,Gross M. Flexible BIOMEMS with electrode arrange:nents on front and back side as key component in neural prostheses and bio- hybrid systems[J]. Sensors and Actuators B, 2002,83 ( 1/2/3 ) : 8 214.
  • 6Rodger D C ,Weiland J D ,Humayun M S,et ol. Scalable high lead- count Parylene package for retinal prostheses[J]. Sensors and Actua- tors B, 2006,117 : 107-114.
  • 7Cheung K C,Renaud P,Tanila H,et ol. Flexible polyimide micro- electrode array for in vivo recordings and current source density analysis[J]. Biosens Bioelectron, 2007,22 : 1 783-1 790.
  • 8Bahhasar C, Patel S, Roy A,et aL Factors affecting perceptual thresh- olds in epiretinal prostheses[J]. Invest Ophthalmol Vis Sci,2008,49 (6) :2 303-2 314.
  • 9Joucla S,Yvert B. Improved focalization of electrical microstimulation using mieroeleetrode arrays:a modeling study [J]. PLoS One,2009, 4(3):2-13.
  • 10Margalit E,Thoreson W B. Inner Retinal Mechanisms Engaged by Retinal Electrical Stimulation[J]. Invest Ophthalmol Vis Sei, 2006,47 (6) :2 606-2 612.