能量代谢守恒法无创血糖检测算法研究 被引量:9

Research on the Algorithm of Non-Invasive Blood Glucose Measurement Based on the Method of Conservation of Energy Metabolism
下载PDF
导出
摘要 根据能量代谢守恒法无创血糖检测的基本原理,设计出无创血糖检测探头并推导出血糖检测的数学模型。采用了多元线性回归和主分量分析方法处理数据,得出了计算血糖值的多元线性表达式,获得了对血糖值贡献最大的两个分量。实验结果显示,采用上述算法计算得出的血糖值与血糖真实值之间的相关系数R=0.86,表明了采用主分量分析和多元线性回归求解无创血糖检测模型的方法是可行的。 According to the theory of non-invasive blood glucose measurement based on the method of conservation of energy metabolism,the sensor for non-invasive blood glucose measurement was designed and the mathematic model for it was derived. By multiple linear regression and principal component analysis, the data were processed, then the multiple linear expression was archived for blood glucose calculation and the two components contributing the most to the blood glucose were found. The experimental results indicated the correlation coefficient R was 0.86 between the test values and the true values of the blood glucose. So the experiments prove that solving the model of non-invasive blood glucose measurement with principal component analysis and multiple linear regression is feasible.
出处 《传感技术学报》 CAS CSCD 北大核心 2013年第7期917-921,共5页 Chinese Journal of Sensors and Actuators
基金 国家863重大项目(2011AA040406) 国家自然科学基金项目(61271119) 国家科技支撑计划(2013BAI03B01) 广西自然科学基金项目(2011jjA40078)
关键词 无创血糖检测 能量代谢守恒法 多元线性回归 主分量分析 non-invasive blood glucose measurement conservation of energy metabolism multiple linear regression principal component analysis
  • 相关文献

参考文献14

  • 1Harutyun M, Emma D, Seungwan K, et al. Non-Invasive in vitro Sensing of D-Glucose in Pig Blood [ J ]. Medical Engineering and Physics ,2012,34 ( 3 ) :299-304.
  • 2Seungwan K, Harutyun M, Jongchel K, et al. Noninvasive in vitro Measurement of Pig-Blood D-Glucose by Using a Microwave Cavity Sensor [ J ]. Diabetes Research and Clinical Practice, 2012,96 ( 3 ) : 379-384.
  • 3Christopher M, David H, Patricia C. Towards Blood Free Measurement of Glucose and Potassium in Humans Using Reverse Iontophoresis [ J ]. Sensors and Actuators B : Chemical, 2012,166 - 167 ( 5 ) :593-600.
  • 4Cai Y Y, Cao D, He X H, et al. Continuous Glucose Monitoring System Based on Smart Phone [ J]. Procedia Engineering, 2012, 29 : 3894-3898.
  • 5Sandeep K V. Non-Invasive Glucose Monitoring Technology in Diabetes Management: A Review [ J ]. Analytica Chimica Acta, 2012,750(10) :16-27.
  • 6Hisako H, Mihara M, Lida T, et al. Blood Glucose Measurement for Flap Monitoring to Salvage Flaps from Venous Thrombosis [ J ]. Journal of Plastic, Reconstructive and Aesthetic Surgery,2012,65 (5) :616-619.
  • 7Dongmin G, David Z, Lei Z, et al. Non-Invasive Blood Glucose Monitoring for Diabetics by Means of Breath Signal Analysis [ J ]. Sensors and Actuators B :Chemical,2012,173 : 106-113.
  • 8Enric M M. Non-Invasive Estimate of Blood Glucose and Blood Pressure from a Photoplethysmograph by Means of Machine Learning Techniques[ J]. Artificial Intelligence in Medicine,2011, 53(2) :127-138.
  • 9Miguel P, Hermann L T, Werner M. Infrared Spectroscopic Analysis of Human Interstitial Fluid in vitro and in vivo Using FT-IR Spectroscopy and Pulsed Quantum Cascade Lasers (QCL): Establishing a New Approach to Non-Invasive Glucose Measurement [ J ]. Spectrochimica Acta Part A : Molecular and Bio- molecular Spectroscopy ,2012,85 ( 1 ) :61-65.
  • 10朱健铭,陈真诚,金星亮,王弟亚.基于DSP技术的便携式无创血糖检测仪[J].电子测量与仪器学报,2009,23(6):108-112. 被引量:13