摘要
目的 探讨有用线束朝向迷路内墙的放射治疗机房的辐射屏蔽计算方法.方法 根据国家职业卫生标准GBZ/T 201.2对某改建机房进行屏蔽计算,并与实际检测结果进行比较.结果 迷路入口处瞬时剂量率的理论计算结果和实际检测结果,防护门内分别为89和86 μSv/h,防护门外分别为5.7和6.2 μSv/h,理论计算结果与实际检测结果具有较好的一致性.结论 通过理论计算结果与检测结果的比较,验证了理论计算方法的准确性.
Objective To study the calculational method for the radiotherapy facilities of the medical linear accelerator' s useful beam towars its maze.Methods The shielding calculation was made under the relevant national standards for a radiotherapy treatment room and compared with the test results.Results The dose rates at the maze entrance as calculated and measured were 89 and 86 μSv/h inside the maze door,as well as 5.7 and 6.2 μSv/h outside the maze door,respectively.The calculated results were consistent with measured results.Conclusions By comparison of calculated results with measured results,the accuracy of the theoretical calculation method could be verified.
出处
《中华放射医学与防护杂志》
CAS
CSCD
北大核心
2013年第4期427-430,共4页
Chinese Journal of Radiological Medicine and Protection
关键词
医用直线加速器
屏蔽
辐射防护
Medical linear accelerator
Shielding
Radiation protection