64排螺旋CT双下肢血管成像低电压扫描的初步研究 被引量:12

Study of 64 Slice Detector CT with Low Tube Voltage in Lower Extremity
原文传递
导出
摘要 目的探讨低管电压(100 kV)扫描下肢CT血管成像(CTA)的辐射剂量、图像质量。方法将60例行双下肢CTA检查的患者,按随机分配原则分为A、B两组(A组30例,B组30例),管电压分别采用100 kV、120kV,管电流均采用300 mA,采用配对t检验统计分析两组下肢CTA的辐射剂量及图像质量。结果扫描结束后记录A组CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP)分别为20.85 mGy、(2342.78±202.49)mGy·cm,B组分别为33.63 mGy、(3478.72±197.34)mGy·cm。A组CTDIvol、DLP较B组分别降低38%;32.6%,两组辐射剂量指标CTDIvol、DLP差异有统计学意义。两组图像质量均能满足诊断要求,两组图像质量评分差异无统计学意义(t=0.521,P=0.604)。结论 64排螺旋CT下肢血管成像采用低管电压可降低辐射剂量的同时获得符合临床诊断要求的图像。 Objective To evaluate the feasibility of lower extremity CT angiography using a low tube voltage ( 100 kV) and the effects of low tube voltage on radiation dose and image quality. Methods 60 patients received lower extremity CT angiography and were randomly assigned to two protocol groups: protocol A (30/60, scanning with 100 kV and 300 mA ) ; protocol B (30/60, scanning with 120 kV and 300 mA ) , CT image quality and radiation does were compared between two groups, T test was performed to analyze the quality. Results CT dose index(group A : 20.85 mGy group B : 33.63 mGy ) and DLP(group A: (2342.78 ± 202. 49 ) mGy · cm group B: (3478.72 ± 197.34)mGy · cm ) were recorded. Compared with protocol B, the CTDIvol and DLP of protocol A were reduced by 38% , 32.6% respectively. There was no significant difference in noise levels among the groups ( P 〉 0.05 ). Conclusion Lower extremity 64 slice CT angiography with 100 KV can reduce radiation dose and obtain a good image quality.
作者 邹平 薛跃君
出处 《临床放射学杂志》 CSCD 北大核心 2013年第9期1340-1342,共3页 Journal of Clinical Radiology
关键词 下肢 体层摄影术 X线计算机 CT血管成像 辐射剂量 Lower extremity Tomography,X-ray computed CT angiography Radiation dosage
  • 相关文献

参考文献6

二级参考文献30

  • 1Naidich DP,Marshall CH,Gribbin C,et al.Low-dose CT of the lungs:preliminary observations.Radiology,1990,175:729-731.
  • 2Madni A,De Maertelaer V,Zanen J,et al.Pulmonary emphysema:radiation dose and section thickness at muhidetector CT quantification:comparison with macroscopic and microscopic morphometry.Radiology,2007,243:250-257.
  • 3Hatayama O,Kobayashi T,Fujimoto K,et al.Utility of single slice high-resolution CT in upper lung field combined with lowdose spiral CT for lung-cancer screening in the detection of emphysema.Intern Med,2007,46:1519-1525.
  • 4de Jong PA,Tiddens HA,Lequin MH,et al.Estimation of the radiation dose from CT in cystic fibrosis Chest,2008,133:1289-1291.
  • 5Donadieu J,Roudier C,Saguintaah M,et al.Estimation of the radiation dose from thorscic CT scans in a cystic fibrosis population.Chest,2007,132:1233-1238.
  • 6高建华,孙宪昶,李剑颖,李娜,夏庆堂,赵雯,戴汝平.不同前置滤线器对64层螺旋CT冠状动脉成像质量及放射剂量影响的对照研究[J].中华放射学杂志,2007,41(8):858-861. 被引量:51
  • 7Schemthaner R, Fleischmann D, Lomoschitz F, et al.Effect of MDCT angiographic findings on the manageement of intermittent claudication.AJR, 2007, 189(5): 1215-1222.
  • 8Fleischmarm D, Hailer RL, Rubin GD. CT angiography of peripheral arterial disease. J Vasc Interv Radiol. 2006 17(1): 3-26.
  • 9Nickoloff EL, Alderson PO.Radiation Exposure to Patients from CT: Reality Perception, and Policy. AJR, 2001, 177(2): 285-287.
  • 10杨智云,陈星,初建平,李树荣,范淼,周旭辉,孟悛非.头颈部多层螺旋CT辐射剂量调查[J].临床放射学杂志,2008,27(5):597-600. 被引量:30

共引文献70

同被引文献90

  • 1Kenneth R.Kase,王恒德.NCRP的辐射防护原则[J].辐射防护通讯,2005,25(2):2-6. 被引量:4
  • 2闵小红,李谭,薛勇利,苑超.64层螺旋CT对下肢动脉疾病的诊断及其临床价值[J].临床放射学杂志,2007,26(8):806-809. 被引量:9
  • 3王佩芬,周康荣,陈祖望,杨培强,沈继章,施惠斌.下肢动脉慢性闭塞症的MRA诊断[J].中华放射学杂志,1997,31(6):396-399. 被引量:17
  • 4Steffens JC, Schafer FK, Oberscheid B, et al. Bolus-chasing Contrast-enhanced 3D MRA of the Lower Extremity. Comparison with intra-arterial DSA[J]. Acta Radiolog- lea, 2003, 44(2): 185-192.
  • 5Janka R, Fellner C, Wenkel E, et al. Con- trast-enhanced MR Angiography of Peri- pheral Arteries including Pedal Vessels at 1.OT: Feasibility study with Dedicated Peripherial angiography Coil[J]. Rdiology, 2005, 235(4): 319-326.
  • 6Yucel EK, Kaufinan JA, Geller SC. Athe- rosclerotic occlusive disease of the lower extremity: prospective evaluation with two-dimensional time-of-flight MR angi- ography[J]. Radiology, 1993,187(3): 637-641.
  • 7Efstathopoulos EP, Kelekis NL, Pantos I, et al. Reduction of the estimated radiation dose and associated patient risk with prospec- tive EGG gated 256 slice CT coronary angiography. Phys Med Brol, 2009,54(17) :5209-5222.
  • 8Frush DP, Donnelly LF, Rosen NS. Computed tomography and radiation risks: Wb.at pediateie health care providers should know. Pediatrics, 2003, 112(4) :951-957.
  • 9Heyer CM, Mohr PS, Lemburg SP, et al. Image quality and ra- diation exposure at pulmonary CT angiography with 100- or 120- kVp protocol: Prospective randomized study. Radiology, 2007, 245(2) : 577-583.
  • 10方光荣.足趾移植拇手指再造与修复的进展[J].中华显微外科杂志,2008,31(3):161-162. 被引量:21

引证文献12

二级引证文献78

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部