比较两种CT模拟定位方法在调强放疗中的误差影响 被引量:5

Errors of two CT simulation positioning methods in intensity-modulated radiotherapy: a comparative study
原文传递
导出
摘要 目的比较调强放疗中CT模拟定位最终等中心标记法和参考点标记法的误差。方法2009--2012年间327例头颈部肿瘤调强放疗患者使用PhilipsBrillianceCTBigBore扫描机及PhilipsTumorLOC模拟定位工作站进行定位,其中最终等中心标记法、参考点标记法定位患者分别为208、119例。在瓦里安Eclipse治疗计划系统进行勾画靶区和设计治疗计划。治疗前使用瓦里安EX直线加速器的机载影像系统千伏级锥形束CT扫描和配准,得出左右、上下、前后方向误差后进行两组比较并成组t检验。最后对5例鼻咽癌患者通过治疗计划系统测量等中心在3个方向移动后对危及器官剂量的影响。结果最终等中心标记法组3个方向平均误差均小于参考点标记法组(P=0.02、0.01、0.03)。等中心在3个方向上移动后的剂量比较结果显示,靶区受量降低的同时靶区周围正常组织受量明显提高,前后方向误差对脊髓和脑干影响最大。结论CT模拟定位最终等中心标记法的误差优于参考点标记法,等中心移动的较小误差可引起危及器官较大的受量变化。 Objective To compare the errors of final isocenter marking method and reference point marking method for CT simulation positioning in intensity - modulated radiotherapy ( IMRT ) . Methods From 2009 to 2012, 327 patients with head and neck cancer for IMRT underwent CT simulation positioning using the Philips Brilliance CT Big Bore scanner and Philips Tumor LOC workstation and were divided into final isocenter marking group ( n = 208 ) and reference point marking group ( n = 119 ) according to positioning methods. Target volume delineation and treatment plan design were performed on the Varian Eclipse treatment planning system (TPS). Before treatment, kilovoltage cone-beam CT scans and registration were performed with the Varian EX on-board imager system to obtain beam position errors in the right-left (RL), superior-inferior (SI), and anterior-posterior (AP) directions, and then comparisons of errors between the two groups were made by independent-samples t test. Finally, the TPS was used to measure the changes in the doses to the organs at risk after moving isocenters in the RL, SI, and AP directions among 5 patients with nasopharyngeal carcinoma. Results The mean beam position errors in the three directions were less in the final isocenter marking group than in the reference point marking group (P = 0. 02,0. 01,0.03 ). After moving isocenters in the three directions, the target dose was reduced and the dose to the normal tissue around the target tumor was increased significantly. The error in the AP direction had the maximum influence on the spinal cord and brainstem. Conclusions Final isocenter marking method leads to less beam position error than reference point marking method in CT simulation positioning. Small isocenter motion can cause large changes in the doses to the organs at risk.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2013年第5期387-390,共4页 Chinese Journal of Radiation Oncology
关键词 CT模拟定位 最终等中心标记法 参考点标记法 头颈部肿瘤 调强放射疗法 CT simulation positioning Final isocenter marking method Reference point marking method Head and neck neoplasms/intensity-modulated radiotherapy
  • 相关文献

参考文献17

  • 1Mutic S, Palta JR, Butker EK, et al. Quality assurance for computed-tomography simulators and the computed-tomography simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66. Med Phys, 2003,30:2762-2792.
  • 2Lee N, Xia P, Quivey JM, et al. Intensity-modulated rediotherapy in the treatment of nasopharyngcal carcinoma: an update of the UCSF experience. Int J Radiot Oncol Binl Phys,2002,53:12-22.
  • 3袁智勇,高黎,徐国镇,易俊林,黄晓东,罗京伟,李素艳,戴建荣.初治鼻咽癌调强放疗的初步结果[J].中华放射肿瘤学杂志,2006,15(4):237-243. 被引量:84
  • 4胡逸民.放射治疗的质量保证与质量控制[A].见:胡逸民 主编.肿瘤放射物理学[C].原子能出版社,1999.614-616.
  • 5钱剑扬.鼻咽癌放射治疗的质量保证和质量控制//夏云飞,钱剑扬,罗伟.实用鼻咽癌放疗学.北京:北京大学医学出版社,2003:151.
  • 6郑步宏,潘建基,徐鹭英,陈传本,张瑜,张秀春.垫内置标记点法测量鼻咽癌适形放疗摆位误差的研究[J].中华放射肿瘤学杂志,2005,14(3):218-221. 被引量:19
  • 7Hunt M. The effect of positional uncertainties on the treatment of primary nasopharynx cancer. Med Phys, 1998,16:456464.
  • 8Stroom JC, de Boer JC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys, 1999, 43:905-919.
  • 9殷蔚伯,余子豪,徐国镇,等.肿瘤放射治疗学(第四版).北京:中国协和医科大学出版社,2007,578-612.
  • 10潘才住,潘建基,陈传本,张秀春,李金莲.应用电子射野影像装置实时纠正鼻咽癌调强放疗摆位误差研究[J].中华放射肿瘤学杂志,2009,18(5):341-345. 被引量:22

二级参考文献119

共引文献383

同被引文献34

引证文献5

二级引证文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部