期刊文献+

原子荧光光谱法测定三氧化二锑中微量铋的测量不确定度评定

Uncertainty evaluation for the determination results of trace bismuth content in the antimony trioxide by atomic fluorescence spectrometry
下载PDF
导出
摘要 文章分析了原子荧光光谱法测定三氧化二锑中微量铋的不确定度来源,对不确定度各分量进行了评定并合成。评定结果表明:测量重复性和待测试液浓度测量是不确定度的主要来源;提高测量准确度的关键是要选择性能可靠的测量设备,提高分析人员的技术水平,严格按标准方法规范操作。 The uncertainty source of determining trace elements Bi in the antimony trioxide by AFS (atomic fluorescence spectrometry) is analyzed in this paper, and each component is evaluated and composed.h turned out that the repeatability of measurement and con centration measurement of the under test liquid are the source of uncertainty. The key to improve accurancy of measurement are choosing the equipments with dependable performance, improving technical level of the analyst, and operating in strict accordance with the standard method.
出处 《企业技术开发》 2013年第9期17-19,33,共4页 Technological Development of Enterprise
关键词 不确定度 三氧化二锑 原子荧光光谱法 uncertainty antimony trioxide bismuth atomic fluorescence spectrometry
  • 相关文献

参考文献4

  • 1GB/T3253.9-2009,锑及三氧化二锑化学分析方法铋量的测定原子荧光光谱法[S].
  • 2JJGl036-2008,国家计量检定规程:电子天平[S] .
  • 3中国实验室国家认可委员会.化学分析中不确定度的评估指南[M].北京:中国计量出版社,2006.
  • 4JJGl96-2006,国家计量检定规程:常用玻璃量器[S].