Biosen-c-clin葡萄糖分析仪的临床应用评价

Clinical evaluation of Biosen-c-clin glucose analyzer application
下载PDF
导出
摘要 目的对Biosen-c-clin葡萄糖分析仪进行临床应用评估,以确定其性能符合临床实验室多中心系统验证。方法随机收集内分泌科患者血液标本,立即使用Biosen-c-clin对血糖进行检测分析:评价仪器的准确性、重复性、线性范围和试剂对血糖的稳定保存时间;评价Biosen-c-clin对血清标本血糖检测与生化仪检测结果的一致性;评价Biosen-c-clin对全血标本血糖检测与罗氏血糖仪检测结果的一致性;评价该仪器多种模式检测血糖间的结果相关性;评价红细胞在全血中所占体积与血糖结果相关性。结果 Biosen-c-clin葡萄糖分析仪准确度偏差小于1%,重复性全血模式高、中、低值的均值和标准差为(13.69±0.039)、(7.08±0.014)、(3.21±0.009);而血清模式为(12.89±0.051)、(6.61±0.031)、(5.87±0.025),本室检测的线性范围相关回归系数为0.995 7,稀释液对血糖稳定保存时间为8h;对血清标本血糖的检测与生化仪所测结果差异有统计学意义(P<0.05),对全血标本血糖的检测与罗氏所测结果比较差异有统计学意义(P<0.05),但均无临床意义;全血葡萄糖随红细胞所占比例的增加而呈减低趋势,同一标本血清葡萄糖为全血葡萄糖的1.13倍。结论 Biosen-c-clin葡萄糖分析仪性能满足临床应用要求。 Objective Clinical assessment of Biosen-c-clin glucose analyzer,to determine its performance meets the clinical laboratory center system verification.Methods To collected randomly blood samples of Department of Endocrinology patients,immediately detected blood glucose by Biosen-c-clin:accuracy,repeatability,linear range of instruments and stable storage time of blood sugar in reagents were evaluated;The consistency between serum glucose detection by Biosen-c-clin and biochemical analyzer was evaluated.The consistency between blood glucose detection by Biosen-c-clin and Roche glucose meter was evaluated.The correlation of glucose detection result by different model was evaluated.The correlation between volume of red cells in the blood and glucose amount was evaluated.Results Accuracy deviation of Biosen-c-clin glucose analyzer less than 1%,Repeated of blood mode:the standard of high,mean and low value difference was(13.69±0.039),(7.08±0.014),(3.21±0.009);and serum mode for(12.89±0.051),(6.61±0.031),(5.87±0.025).The regression coefficient of the linear range was 0.995 7.Stable storage time of blood sugar in reagents was 8hours.Comparison of glucose detected in different instruments have Statistical significance(P〈0.05),but had no clinical significance.Blood glucose will decrease with red cell proportion increasing,the same specimens of serum glucose was 1.13times of blood glucose.Conclusion The performance of Biosen-c-clin glucose analyzer meet the requirement of clinical application.
出处 《国际检验医学杂志》 CAS 2013年第18期2413-2414,2416,共3页 International Journal of Laboratory Medicine
关键词 血糖 临床应用 实验室技术和方法 blood glucose clinical application laboratory techniques and procedures
  • 相关文献

参考文献10

二级参考文献20

  • 1American Diabetes Association. Consensus statement on self-monitoring of blood glucose [ J ]. Diabetes Care, 1987,10( 1 ) :95-99.
  • 2American Diabetes Association. Bedside blood glucose monitoring in hospitals [ J ]. Diabetes Care, 2004, 27 (Suppl 1 ) :S104.
  • 3G13/T19634-2005.体外诊断检验系统自测用血糖监测系统通用技术条件[S].
  • 4Bastanhagh MH, Shirvan AR, Heshmat R, et al. Evaluation of the efficacy of blood glucose home monitoring devices [ J]. Med Sci Monit, 2007, 13 (3) :PI1-PI6.
  • 5Skeie S, Thue G, Nerhus K, et al. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician [J]. Clin Chem, 2002, 48(7) : 994-1003.
  • 6Solnica B, Naskalski JW, Sieradzki J. Analytical performance of glucometers used for routine glucose self- monitoring of diabetic patients [ J ]. Clin Chim Acta, 2003,331 (1-2) :29-35.
  • 7Stahl M, Brandslund I. Measurement of glucose content in plasma from capillary blood in diagnosis of diabetes mellitus[J]. Scand J Clin Lab Invest ,2003, 63(6) : 431-440.
  • 8Joint Commission on Accreditation of Healthcare Organization(JACHO).Champion of quality in health care.Chicago:Greenwich Publishing Groups,2001.
  • 9Burner D.Understanding accreditation in laboratory medicine.London:Piggott Printers Ltd,1996:4-11.
  • 10Joint Commission International.Joint Commission international accreditation standards for hospitals.2nd.Chicago:Jiont Commission Resurces,2002.

共引文献164

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部