期刊文献+

纯钛在细胞培养液中的腐蚀性能研究 被引量:1

Study of corrosion behavior of titanium in cell culture medium
下载PDF
导出
摘要 目的:纯钛在细胞培养液中的腐蚀性能研究。方法:腐蚀介质采用Hank's液为对照组,细胞培养液DMEM为实验组。利用电化学工作站测定钛试件在电解质中的极化曲线和电化学阻抗谱,通过PowerSuite软件获得腐蚀参数,对比研究高仿真生理环境对纯钛腐蚀性能的影响。结果:纯钛试件在对照组和实验组的自腐蚀电流密度分别为(84.2±12.3)mA/cm2和(49.8±7.40)mA/cm2(P<0.05)。电化学阻抗谱分析获得对照组和实验组的阻抗值分别为(72.5±10.1)×104Ω和(253.1±22.6)×104Ω(P<0.05)。结论:生理环境中的蛋白成分可能影响到钛种植材料腐蚀性能,提高其抗腐蚀性能。 Objective: To investigate the corrosion resistance of the titanium in cell culture medium. Methods: The electrochemical behavior of the titanium in Hankg solution and cell culture medium DMEM was studied by electrical impedance spectroscopy (EIS) and potentiodynamic polarization test using electrochemical methods. Data were analysed by PowerSuite software. Results: The self- corrosion current of the titanium in Hankg solution and DMEM was ( 84.2 ± 12.3 ) mA/em2 and (49.8 ± 7.4 ) mA/cm2 respective- ly. The titanium corrosion resistance in Hankg solution and DMEM was (72.5 ± 10.1) Ω ×104 and (253.1± 22.6 ) Ω ×104 re- spectively. Conclusion: The protein in physiological environment may play an important role in corrosion behavior of titanium.
作者 刘锁钢 刘莉
出处 《实用口腔医学杂志》 CAS CSCD 北大核心 2013年第5期630-633,共4页 Journal of Practical Stomatology
关键词 纯钛 腐蚀性能 生理环境 Titanium Corrosion Physiological environment
  • 相关文献

参考文献9

  • 1钱棱,王培志.种植体腐蚀及其影响因素的研究进展[J].口腔生物医学,2012,3(2):87-89. 被引量:4
  • 2Zaffe D, Bertoldi C, Consolo U. Element release from titanium devices used in oral and maxillofacial surgery[ J]. Biomateri- als, 2003,24(6) : 1093 - 1099.
  • 3Messer RL, Tackas G, Miekalonis J, et al. Corrosion of ma- chined titanium dental implants under inflammatory condi- tions[J]. J Biomed Mater Res B Appl Biomater, 2009, 88 (2) :474 -481.
  • 4Yu WQ, Qiu J, Zhang FQ. In vitro corrosion study of differ- ent TiO2 nanotube layers on titanium in solution with serum proteins [ J ]. Colloids Surf B Biointerfaees, 2011,84 ( 2 ) : 400 - 405.
  • 5Karpagavalli R, Zhou A, Chellamuthu P, et al. Corrosion behavior and biocompatibility of nanostructured TiO2 film on Ti6A14V[J]. J Biomed Mater Res A, 2007,83(4) :1087 - 1095.
  • 6Kumar S, Narayanan TS. Corrosion behaviour of Ti- 15Mo al- loy for dental implant applications [ J ]J Dent, 2008,36(7) :500 - 507.
  • 7Takemoto S, Hattori M, Yoshinari M, et al. Corrosion be- havior and surface characterization of titanium in solution containing fluoride and albumin[ J]. Biomaterials, 2005,26 (8) :829 -837.
  • 8Huang HH. Effects of fluoride and albumin concentration on the corrosion behavior of Ti-6A1-4V alloy[ J ]. Biomaterials,2003, 24(2) :275 -282.
  • 9陈建治,彭红艳,王磊,石玉龙.氟离子对纯钛表面微弧氧化膜耐腐蚀性能的影响[J].实用口腔医学杂志,2012,28(2):164-167. 被引量:5

二级参考文献27

共引文献7

同被引文献1

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...