基于CTVision研究宫颈癌调强放疗中摆位误差对剂量分布的影响 被引量:10

Study of the Influence of Set-up Errors on Dose Distribution of Intensity-modulated Radiation Therapy for Cervical Cancer Based on CTVision
下载PDF
导出
摘要 目的通过测量宫颈癌调强放疗中的摆位误差,分析摆位误差对肿瘤计划靶体积(PTV)及危及器官受照剂量分布的影响。方法 20例宫颈癌患者调强放疗过程中每周1次行CTVision图像引导扫描,其实时扫描图像与计划CT图像融合配准获得X、Y、Z方向上的摆位误差,将上述误差引入治疗计划系统重新计算得到PTV及危及器官的剂量分布参数并与原计划行配对t检验。结果X、Y、Z方向上的摆位误差分别为2.14、5.52、2.47 mm;再计划的PTV 95%体积剂量(D_(95))和最小剂量(Dmin)明显小于原计划(P<0.05);再计划膀胱和直肠接受45 Gy剂量体积(V_(45))明显大于原计划(P<0.05)。结论摆位误差导致PTV D_(95)明显不足,危及器官体积剂量均增加,因此宫颈癌放疗过程中摆位误差的在线纠正是必要的。 Objective To analyze the influence of set-up errors on dose distribution of tumor planning target volume (PTV) and organs at risk by measuring the set-up errors of intensity-modulated radiation therapy for cervical cancer. Methods CTVision image guided scanning of 20 patients with cervical cancer was conducted once a week with a total of 5 times during IMRT. The set-up errors on X axis, Y axis and Z axis were obtained through the registration of real-time scanning images and planning CT images of CTVision image guided scanning. The dose distribution parameters of PTV and organs at risk were obtained based on the calculation of set-up errors data with treatment planning system. Then comparing these parameters with those of original plan by paired t-test. Results The average values of set-up errors on X axis, Y axis and Z axis were 2.14, 5.52, 2.47 mm, respectively. The PTV 95% volume dose (D95) and PTV minimal dose of the replanning were lower than those of the original plan (P 〈0.05). The 45Gy dose volume (V45) received by bladder and rectum of the replanning were higher than those of the original plan(P 〈0.05). Conclusion The obviously insufficient dose of PTV D95 and increased dose of V45 of the bladder and rectum were caused by setup errors, which indicated that the online adjustment of the setup errors is necessary during the intensity-modulated radiation therapy for cervical cancer.
出处 《中国医疗设备》 2013年第9期112-114,共3页 China Medical Devices
基金 广西科学研究与技术开发计划课题(2011D39)
关键词 CTVision图像引导放疗系统 宫颈癌 调强放疗 摆位误差 计划靶体积 临床靶体积 CTVision image guided radiation therapy cervical cancer IMRT set-up errors planningtarget volume clinical target volume
  • 相关文献

参考文献9

  • 1Mell LK,Roeske JC,Mundt AJ,et al.A survey of intensity-modulated radiation therapy use in the United States[J].Cancer,2003,98(1):204-211.
  • 2Beadle BM,Jhinggran A,Salehpour M,et al.Cervix regression and motion during the comse of external beam chemoradiation for cervical cancer[J].IntJ Radiat Oncol Biol Phys,2009,73(1):235-241.
  • 3姜书梅,王仁本.盆腔放疗中放射性直肠损伤的预防[J].中华肿瘤防治杂志,2008,15(3):235-238. 被引量:2
  • 4Van de Bunt L,Van der Heide UA,Ketehars M,et al.Conventional, conformal,and intensity-modulated radiation therapy treatment planning of external beam radiotherapy for cervical cancer.the impact of tumor regression[J].Int J Radiat Oncol Biol Phys,2006,64(1):189-196.
  • 5Shimizu S,Shirato H,Kagei K,et al.Impact of repiratory movement in the computed tomographic images of small lung tumors in three-dimensional (3D) radiotherapy[J].Int J Radiat Oncol Biol Phys,2000,46(5):1127-1133.
  • 6王诗琴,张书旭,余辉,谭剑明.鼻咽癌进行图像引导放疗时不同配准方法对体位校正的影响[J].中国医疗设备,2010,25(4):77-79. 被引量:6
  • 7戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 8Wilder R B,Chittenden L,Msea A V,et al.A prospective study of intraffaction prostate motion in the prone vs.supine position[J].Int J Radiat Oncol Biol Phys,2010,77(1):165-170.
  • 9庞皓文,邱杰,孙帅,杨波,管秋,张福泉.宫颈癌调强放疗中直肠体积与位置变化及对吸收剂量影响[J].中华放射肿瘤学杂志,2012,21(3):258-261. 被引量:31

二级参考文献53

  • 1谢楚平,叶伟坤,林关朋,张萍.直肠癌术后局部复发的诊断和治疗[J].大肠肛门病外科杂志,2005,11(2):99-101. 被引量:3
  • 2朱正飞,徐志勇,傅小龙.电子射野影像系统临床应用的研究进展[J].中国癌症杂志,2006,16(2):154-157. 被引量:7
  • 3张照辉,贾廷珍,沈彤,王俊杰,张淑兰,马力文.放射性直肠炎的诊断和治疗——附97例病例分析[J].中华放射医学与防护杂志,2006,26(1):29-30. 被引量:47
  • 4于金明,袁双虎.图像引导放射治疗研究及其发展[J].中华肿瘤杂志,2006,28(2):81-83. 被引量:99
  • 5戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 6Mell LK, Roeske JC, Mundt A J, et al. A survey of intensity modulated radiation therapy use in the United States. Cancer, 2003,98:204-211.
  • 7Roeske JC, Lujan A, Rotmensch J, et al. Intensity-modulated whole pelvic radiation therapy in patients with gynecologic malignancies. Int J Radiat Oncol Biol Phys,2000,48:1613-1621.
  • 8Mell LK, Tiryaki H, Ahn KH, et al. Dosimetric comparison of bone marrow-sparing intensity-modulated radiotherapy versus conventional techniques for treatment of cervical cancer. Int J Radiat Oncol Biol Phys ,2008,71 : 1504-1510.
  • 9Chen MF, Tseng CJ, Tseng CC, et al. Clinical outcome in post hysterectomy cervical cancer patients treated with concurrent Cisplatin and intensity-modulated pelvic radiotherapy: comparison with conventional radiotherapy. Int J Radiat Oncol Bio! Phys, 2007,67 : 1438-1444.
  • 10Beadle BM, Jhingran A, Salehpour M, et al. Cervix regression and motion during the course of external beam ehemoradiation for cervical cancer. Int J Radiat Oncol Biol Phys,2009,73:235-241.

共引文献217

同被引文献102

引证文献10

二级引证文献39

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部