期刊文献+

作业人员尿氟测定的不确定度评定

Evaluation of uncertainty in content determination of urinary fluoride of workers
原文传递
导出
摘要 目的对作业人员尿氟测定进行不确定度评定。方法通过分析尿中氟的离子选择电极测定方法,找出引起测量不确定度的来源并对其进行量化。结果用本方法测定作业人员尿氟的扩展不确定度为0.020mg/L。结论尿氟测定的测量不确定度校准曲线拟合贡献最大,其次是标准物质本身和标准系列配制的贡献较大。 Objective To evaluate the uncertainty in content determination of urinary fluoride of workers.Methods The electrode method was chosen by analyzing urinary fluoride ion.The source of the uncertainty was found and quantified.Results The expanded uncertainty was 0.020mg / L by using this method for the determination of urinary fluoride of workers.Conclusions The contribution of calibration fitting curve was the largest,followed by the standard material itself and the standard series.
出处 《医学动物防制》 2013年第9期1058-1059,1061,共3页 Journal of Medical Pest Control
关键词 作业人员 尿氟 测量不确定度 Workers Urine fluoride Uncertainty of measurement
  • 相关文献

参考文献3

  • 1国家技术质量监督局.JJF1059.1-2012测量不确定度评定与表示.北京:中国计量出版社,2012.
  • 2中华人民共和国卫生部.WS/T30-1996尿中氟的离子选择电极测定方法.北京:中国标准出版社,1996.
  • 3国家质量监督检验检疫总局.JJG196-2006常用玻璃量器检定规程.北京:中国计量出版社,2006.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部