Visante-OCT、Sirius及A超测量中央角膜厚度的比较 被引量:3

Comparison of Central Corneal Thickness Measurements by Visante-OCT,Sirius and A-scan
下载PDF
导出
摘要 目的比较Visante-OCT、Sirius与A超测量中央角膜厚度(central corneal thickness,CCT)的差异性,评价三种仪器测量结果的一致性。方法分别用Visante-OCT、Sirius和A超测量43例屈光不正志愿者(均为右眼)的CCT,数据行单因素方差分析、简单线性相关及Bland-Altman分析。结果 Visante-OCT、Sirius和A超测量的CCT分别为(522.88±33.53)μm、(538.35±35.05)μm和(532.23±34.16)μm;三种仪器测量结果两两之间均有高度相关性;A超与Visante-OCT测量结果无统计学差异(P=0.208),测量差值为(9.35±7.39)μm,两者95%的一致性区间为(-3.4~22.1)μm;A超与Sirius测量CCT差异无统计学意义(P=0.409),测量差值为(-6.12±7.39)μm,两者95%的一致性区间为(-18.9~6.7)μm;Visante-OCT与Sirius测量CCT差异具有统计学意义(P=0.038),测量差值为(-15.47±7.39)μm,两者95%的一致性区间为(-30.2^-0.7)μm。结论 Visante-OCT与Sirius用于测量CCT,与A超的一致性较好,在临床使用中可替代A超,但需谨慎选择;Visante-OCT与Sirius测量CCT存在一定差异,两者不能替换使用。 Objective To evaluate the differences and agreement in central corneal thickness (CCT) measurements by Visante-OCT, Sirius and A-scan. Methods The CCTs of 43 healthy eyes were measured by Visante-OCT, Sirius and A-scan. The differences, correlations and agreements among three measurements were evaluated with one way analysis of variance (ANOVA), Pearson correlation coefficients and Bland-Altman plots respectively. Results The measured CCTs were (522. 88 ± 33.53)μm, (538. 35 ± 35.05)μm and (532. 23 ± 34. 16) μm respectively by Visante-OCT, Sirius and scan. There existed high pairwise correlations. The difference between A-scan and VisanteOCT was (9. 35 ± 7. 39)μm which was not significant (P = 0. 208), and the 95% limit of agreement was ( -3.4 - 22. 1 )μm. The difference between A-scan and Sirius was ( - 6. 12 ± 7. 39)μm which was not significant (P = 0. 409), and the 95% limit of agreement was ( - 18. 9 - 6. 7)μm. The difference between Visante-OCT and Sirius measurements was ( - 15.47 ± 7. 39)μm which was significant (P = 0. 038), and the 95 % limit of agreement was( - 30.2 - - 0. 7)μm. Conclusion Visante-OCT and Sirius have good coincidence with A-scan in measuring CCT and can replace A-scan in clinic, but cautious selection is needed. Visante-OCT and Siriu measurements have difference and can't replace each other.
出处 《华南国防医学杂志》 CAS 2013年第9期630-633,共4页 Military Medical Journal of South China
关键词 Visante-OCT Sirius A超 中央角膜厚度 一致性 Visante-OCT Sirius A-scan Central corneal thickness Agreement
  • 相关文献

参考文献15

  • 1Amano S, Honda N, Amano Y, et al. Comparison of central cor neal thickness measurements by rotating Scheimpflug camera, ul- trasonic pachymetry, and scanning- slit corneal topography[J]. Ophthalmology, 2006,113 (6) : 937-941.
  • 2Pakravan M,Parsa A,Sanagou M,et al. Central corneal thickness and correlation to optic disc size: a potential link for susceptibility to glaucoma[J]. Br J Ophthalmol,2007,91 (1) :26 28.
  • 3Romero-Jim6nez M, Santodomingo-Rubido J, Wolffsohn JS. Kera toeonus: a review[J]. Cont Lens Anterior Eye, 2010, 33 ( 4 ) : 157- 166.
  • 4Doors M, Cruysberg LP, Berendschot TT, et al. Comparison of central corneal thickness and anterior chamber depth measure ments using three imaging technologies in normal eyes and afterphakic intraocular lens implantation[J], Grades Arch Clin Exp Ophthalmol, 2009,247(8) : 1139-1146.
  • 5何燕玲,黎晓新,鲍永珍,刘国栋,胡亦文.Pentacam系统与A超角膜测厚仪测量瞳孔中心角膜厚度的比较[J].中华眼科杂志,2006,42(11):985-988. 被引量:28
  • 6Bland JM, Altman 13(3. Statistical methods for assessing agree- ment between two methods of clinical measurement[J]. Lancet, 1986,1 (8476) 307-310.
  • 7Zhao PS,Wong TY, Wong WL, et al. Comparison of central cor- neal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry [J]. Am J Ophthalmol, 2007,143 (6) : 1047-1049.
  • 8Kim HY, Budenz DL, Lee PS, et al. Comparison of central corneal thickness using anterior segment optical coherence tomography vs. ultrasound pachymetry[J]. Am J Ophthalmol,2008,145(2) : 228-232.
  • 9Li Y, Shekhar R, Huang D. Corneal paehymetry mapping with high-speed optical coherence tomography [J]. Ophthalmology, 2006,113 (5) : 792-799.
  • 10江先明,杨斌,王铮,黄国富,罗晓阳,张晓晓,周胜.四种仪器测量中央角膜厚度的比较[J].中国实用眼科杂志,2009,27(5):467-471. 被引量:16

二级参考文献69

  • 1赵玲,黄俊芝.正常人与原发性闭角型青光眼患者周边前房深度的分析[J].中国煤炭工业医学杂志,2004,7(9):812-813. 被引量:4
  • 2杨斌,王铮,黄国富,张醇,吴君舒,禤俭环,李春满.中国近视患者的角膜厚度与屈光度关系的研究[J].中国实用眼科杂志,2004,22(7):516-518. 被引量:18
  • 3李坤,陆蓓,崇晓霞,陈彦玲,李晓玲.麻醉剂和泪膜对健康人角膜厚度的影响[J].眼视光学杂志,2004,6(4):244-246. 被引量:16
  • 4何燕玲,黎晓新,鲍永珍,刘国栋,胡亦文.Pentacam系统与A超角膜测厚仪测量瞳孔中心角膜厚度的比较[J].中华眼科杂志,2006,42(11):985-988. 被引量:28
  • 5Lackner, B.Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and uhrasound.Optom Vis Sci, 2005,82(10) : 892-9.
  • 6Hashemi,H.and S.Mehravaran,Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia.J Cataract Refract Surg, 2007,33(10) :p.1701-7.
  • 7Prisant, O.Reliability of pachymetric measurements using orbscan after excimer refractive surgery.Ophthalmology, 2003,110(3 ) : 511-5.
  • 8Lu Sanetis,U.Reproducibility and repeatability of central corneal thickness measurement in keratoconus using the rotating Seheimpflug camera and ultrasound pachymetry.Am J Ophthalmol, 2007,144 (5): 712-718.
  • 9Zhao, P.S.Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetr.Am J Ophthalmol, 2007,143 (6) :1047-9.
  • 10Li,Y.,R.Shekhar,and D.Huang,Corneal pachymetry mapping with high -speed optical coherence tomography.Ophthalmology, 2006,113 (5):792-9.

共引文献68

同被引文献29

引证文献3

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部