摘要
对流动注射法测定地表水中高锰酸盐指数进行了深入研究,优化了仪器的最佳工作参数。其测定结果表明,该方法在标准溶液质量浓度为0~10.8mg/L时工作曲线具有良好的线性,最低检测限为0.0748mg/L,相对标准偏差为0.59%-4.70%,加入标准物质回收率为98.5%。101%。该方法与酸性高锰酸钾法进行比对,其测定结果准确、可靠,无显著性差异。
This paper studied a method that determines permanganate index in the surface water using flow injec- tion analysis( FIA), and optimized the optimal working conditions of facilities. Determination results show that the line- arity range for the determination of mercury was found to be 0 - 10.8 mg/L. The lower detection limit was found to be 0.0748 mg,/L,RSD obtained was 0.59 % -4.70 % and with addition of standard substance the recovery was in the range of 98.5 -101%. Compared to acidic permanganate,the test results are precise and reliable,having no obvious inconsistency.
出处
《黄金》
CAS
2013年第10期76-79,共4页
Gold
关键词
流动注射法
高锰酸盐指数
国家标准方法
方法比较
flow injection analysis
permanganate index
the national standard method
methods comparison