期刊文献+

颈椎椎间隙测量及同种异体骨颈椎融合骨垫的研制

Measure of Chinese people cervical vertebral interspaces and the design of allograft bone cervical fusion spacer
下载PDF
导出
摘要 目的根据MRI测量正常国人颈椎椎间隙数据,研制符合国人解剖学特征的同种异体骨颈椎椎间融合骨垫。方法通过MRI测量100例C2,3至C6,7每个椎间隙的前缘、中点、后缘高度值及横径宽度值、矢状径长度值,其中男性有50人(20~55岁,平均年龄38.7岁),女性有50人(20~55岁,平均年龄36.1岁)。并根据该数据设计同种异体骨颈椎椎间融合骨垫。结果同节段颈椎椎间隙正中矢状面后缘高度<前缘高度<中间高度(P<0.05),椎间隙横径大于矢状径(P<0.05),同节段同部位的测量值,男性组大于女性组(P<0.05);同种异体骨颈椎椎间融合骨垫设计为前缘高度为6.0、7.0、8.0、9.0 mm 4个型号的楔状,其表面具有倒齿结构,中央有脱钙骨基质(DBM)置入孔,用于置入DBM粉末。结论依据MRI测量的颈椎椎间隙数据设计的同种异体骨颈椎融合骨垫符合国人解剖学要求。 Objoetive Based on the M IRI measure values of cervical vertebral interspace of healthy Chinese people, we aim to design the cervical allograft bone spacer in anterior cervical fusion to match the anatomical characteristics of Chinese people. M^aoda One hundred young normal Chinese people (50 males, aged 20-55 years,with an average of 38.7 years; 50 females, aged 20-55 years,with an average 36.1 years) were scanned by MRI according to gender. Measurements of the transverse diameter and sagittal diameter, as well as the anterior edge height, midpoint height and posterior edge height, were surveyed with C2,3NC6,7 vertebral interspaces. According to the data, we designed the allograft bone cervical fusion spacer. Results Located in the midline sagittal, the posterior edge height of cervical intervertebral space at the same segment was shorter than the leading edge height, which was further shorter than midpoint height (P〈0.05). The transverse diameter of cervical intervertebral space was greater than the sagittal diameter (P〈0.05). The male's measure values of the same segment was greater than the female's (P〈0.05). The cervical allograft bone spacer was designed wedge-shaped and had 6,7,8,9 mm four models according to its anterior edge height. There had inverted tooth structure on the surface. The demineralized bone matrix (DBM) placed hole, which was filled with the power of DBM, was located in the center of spacer. Conclusion The cervical allograft bone spacer which was designed with the MIRI measure values of cervical vertebral interspace was compliant with the anatomy data of Chinese people.
出处 《中国临床解剖学杂志》 CSCD 北大核心 2013年第5期511-513,517,共4页 Chinese Journal of Clinical Anatomy
基金 武汉市科技攻关项目(201161038344-05)
关键词 颈椎椎间隙 解剖 同种异体骨 骨垫 脱钙骨基质 Cervical vertebral interspace Anatomy Allografl bone Spacer Demineralized bonematrix (DBM)
  • 相关文献

参考文献12

  • 1Theo HS, Ralph M, Martin VD, et al. Change in bone architecture during spinal fusion:three years follow-up and the role ofcage stiffness [J]. Spine, 2003, 28(16): 1802-1809.
  • 2Kansyama M, Hashimoto T, Shigenobu K, et al. Pitfalls of anterior cervical fusion using titanium mesh and local autogratt [J]. Spinal Disord Teeh, 2003, 16(6): 513-518.
  • 3Cahill DW, Martin GJ, Hajjar MV, et al. Suitability of bioresorbable cages for anterior cervical fusion [J]. J Neurosurg, 2003, 98 (2): 195 -201.
  • 4Lu j, Ebraheim NA, Yang H, et al. Anatomic bases for anterior spinal surgery: surgical anatomy of the cervical vertebral body and disc space [J]. Surg Radiol A.nat, 1999, 21(4): 235-239.
  • 5谢兴国,许仕全,代小思,刘尚清.颈椎间隙的应用解剖学研究[J].四川解剖学杂志,2000,8(4):203-205. 被引量:6
  • 6张亘瑷,江建明,金大地.人工颈椎间盘置换术的解剖学参数测量[J].中国临床解剖学杂志,2006,24(3):283-285. 被引量:7
  • 7王刚,陈开润.颈椎椎体间隙的应用解剖学[J].解剖学杂志,2010(3):381-382. 被引量:3
  • 8张旗.同种骨移植与骨库的历史与现状[J].生物骨科材料与临床研究,2004,1(5):46-48. 被引量:2
  • 9Kandziora F, Pnugmacher R, schafer J, et al. Biomechanical comparisonof cervical spine interbody fusion cages [J]. Spine, 2001, 26 (17): 1850-1857.
  • 10宋文慧,李宝兴,马迅,周劲松,王岩.颈椎同种异体骨笼椎间融合的组织学研究[J].中国临床解剖学杂志,2006,24(5):557-559. 被引量:4

二级参考文献34

  • 1王岩,肖嵩华,陆宁,张雪松.颈人工椎间盘假体置换术的临床应用[J].中华外科杂志,2004,42(21):1333-1337. 被引量:34
  • 2夏虹,刘景发,吴增晖,朱青安,钟世镇,原林.腰椎间盘突出症不同术式的三维运动范围研究[J].中国临床解剖学杂志,1995,13(1):56-57. 被引量:13
  • 3张亘瑷,江建明,金大地.人工颈椎间盘置换术的解剖学参数测量[J].中国临床解剖学杂志,2006,24(3):283-285. 被引量:7
  • 4Martin V D,Theo H S,Sugihara S,et al.The effect of cage stiffness on the rate of lumbar interbody fusion:an in vivo model using poly and titanium cages[J].Spine,2002,27 (2)..682-688.
  • 5Theo H S,Ralph M,Martin V D,et al.Change in bone architecture during spinal fusion:three years follow-up and the role of cage stiffness[J].Spine,2003,28(16):1802-1809.
  • 6Kanayama M,Hashimoto T,Shigenobu K,et al.Pitfalls of anterior cervical fusion using titanium mesh and local autograft[J].Spinal Disord Tech,2003,16(6):513-518.
  • 7Cahill D W,Martin G J,Hajjar M V,et al.Suitability of bioresorbable cages for anterior cervical fusion[J].J Neurosurg,2003,98(2):195-201.
  • 8Lu J,Ebraheim N A,Yang H,et al.Anatomic bases for anterior spinal surgery:surgical anatomy of the cervical vertebral body and disc space[J].Surg Radiol Anat,1999,21(4):235-239.
  • 9Tanaka M,Nakahara S,Inoue H.A pathologic study of discs in the elderly separation between the cartilaginous endplate and the vertebral body[J].Spine,1993,18(11):1456-1462.
  • 10Pait T G,Killefer J A,Arnautovic K I.Surgical anatomy of the anterior cervical spine:the disc space,vertebral artery,and associated bony structures[J].Neurosurgery,1996,39(4):769-776.

共引文献14