不同条件下Propex全自动根尖定位仪定位水平斜行根折位置准确性的体外研究

In vitro evaluation of the accuracy of Propex electronic apex locators in positions of simulated horizontal oblique root fractures under various conditions
下载PDF
导出
摘要 目的采用体外模型研究Propex全自动根尖定位仪在不同环境下定位水平斜行根折位置的准确性。方法选取30颗离体前磨牙,采用金刚砂车针从牙根中下段颊侧面以65°的角度斜向冠方破坏牙根,根管暴露即停止切磨且保证根管断面距根尖8mm,建立水平斜行根折体外模型。然后用Propex根尖定位仪分别于0.9%生理盐水、3%双氧水、17%乙二胺四乙酸(EDTA)3种条件下测量根折位置。对测量结果采用χ2检验进行统计学分析。结果 Propex全自动根尖定位仪在0.9%生理盐水、3%双氧水、17%EDTA条件下测得的准确率分别是83.33%,86.67%,80.00%,虽然3种液体环境下测得的准确率不同,但这种差异没有统计意义(P>0.05)。结论 Propex全自动根尖定位仪能够比较准确定位水平斜行根折的位置,且不受液体环境影响。 Objective To evaluate the accuracy of Propex electronic apex locator in positions of simulated horizontal oblique root fractures under various conditions.Methods Thirty premolar teeth with simulated horizontal oblique root fractures were selected,and simulated horizontal incomplete root fracture through the preparation of an oblique notch on the buccal root surface approximately 65°with respect to the root axis.The notch was 8mm from the anatomic apex of the root,completely exposing the root canal.Propex electronic apex locator was applied to determine the fractures plane of the canals being flooded with three different solutions.The working lengths were also visually measured and compared with the electronic reading by usingχ2 test.Results While the canals were flooded with 0.9% NaCl,3% hydrogen peroxide and 17% EDTA,the ratios of the distance within limits of±1.00mm were 83.33%,86.67%and 80.00%,respectively.There was no significant difference among the three different solutions(P 0.05).Conclusion Propex could measure the length accurately of coronal root canal segment in teeth,simulated horizontal oblique root fractures under different solutions and without any influencing.
出处 《检验医学与临床》 CAS 2013年第19期2558-2559,共2页 Laboratory Medicine and Clinic
关键词 Propex全自动根尖定位仪 水平斜行根折 液体 工作长度 Propex simulated horizontal oblique root fracture solution working length
  • 相关文献

参考文献8

  • 1imcena-Martin C,Robles-Gij6n V,Ferrer-Luque CM,et al. In vitro evaluation of the accuracy of three electronic apex loca- tors[J]. J Endod, 2004,30(4) : 231-233.
  • 2Herrera M, Abalos C,Planas A J, et al. Influence of apical constriction diameter on Root ZX apex locator precision [J]. J Endod,2007,33(8) :995-998.
  • 3Briseno-Marroquin B, Frajlich S, Goldberg F, et al. Influ ence of instrument size on the accuracy of different apex locators:an in vitro study[J]. J Endod, 2008,34(6) : 698 702.
  • 4Aurelio JA, Nahmias Y, Gerstein H. A model for demon- strating an electronic canal length measuring device[J]. J Endod, 1983,9(12) : 568-569.
  • 5Goldberg F,Frajlich S, Kuttler S, et al. The evaluation of four electronic apex locators in teeth with simulated hori- zontal oblique root fractures [J]. J Endod, 2008,34 (12) 1497-1499.
  • 6Cvek M, Mejare I, Andreasen JO. Conservative endodontic treatment of teeth fractured in the middle or apical part o the root[J]. Dent Traumatol,2004,20(5) :226-261.
  • 7Goldberg F,De Silvio AC, Manfr6 S, et al. In vitro meas urement accuracy of an electronic apex locator in teeth with simulated apical root resorption[J]. J Endod, 2002, 28(6) :461-463.
  • 8宋卫健.根管长度测量仪测定前牙根管工作长度准确性的评价[J].口腔医学,2007,27(1):41-42. 被引量:4

二级参考文献5

  • 1凌均棨.牙髓病学[M].北京:人民卫生出版社,1998.118-119.
  • 2Recard O,Roux D,Woda A.Clinical evaluation of the accuracy of the Evident RCM Mark II apex locator[J].J Endodon.1991,17(11):567-569.
  • 3Kelleme ME,Brown CEJr,Newton CW.A clinical evaluation of the endocater-an electronic apex locator[J].J Endodon,1991,17(6):271-274
  • 4Shanhrokh S,William WY,Gluskin AH.An in vivo evaluationof root ZX electronic apex locator[J].J Endodon.1996,22(11):616-618.
  • 5Macdonald NJ,Hovland EJ.An evaluation of the apex locator endodenter[J].J Endodon,1990,16(1):5-8.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部