颈动脉CTA低管电压扫描研究 被引量:3

The Carotid Artery CTA Scan Study in Lower Tube Voltage
下载PDF
导出
摘要 目的:探讨颈动脉CTA扫描时通过降低管电压以达到降低辐射剂量的可行性。方法:将40例需行颈动脉CTA检查的患者依据管电压不同而分为A(120 kV)、B(80 kV)两组,每组20例。分析比较两组图像的CT值、对比噪声比(CNR)、信号噪声比(SNR)及辐射剂量,并对重建图像质量、血管边缘形态进行评估分析。结果:B组容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP)及有效剂量(ED)较A组分别降低63.97%、63.83%及63.79%,图像质量评价结果:B组颈动脉平均CT值较A组显著增加,差异具有统计学意义(P<0.05)。A、B两组CNR、SNR值组间比较及两两比较差异均无显著统计学意义(P>0.05);且A、B两组颈动脉重建图像质量、血管边缘形态整体满意,无明显差别(P>0.05)。结论:80 kV低管电压颈动脉CTA扫描在不影响图像质量的前提下,最大能够降低约64%的辐射剂量。因此,颈动脉CTA检查通过降低管电压的低剂量扫描是可行的,可常规使用。 Objeetlve: To explore the carotid artery CTA scan through lower tube voltage in order to achieve the feasibility of reducing radiation dose. Method : 40 patients of required carotid artery CTA examination were divided into two groups : according to the tube voltage A ( 120 kV ), B ( 80 kV ), 20 cases in each group. Analysed and compared two sets of image CT value, contrast to noise ratio ( CNR ), signal to noise ratio ( SNR ) and radiation dose, meantime evaluated the reconstructed image quality, vascular morphology. Result: The CT dose index ( CTDIvol ), dose length product ( DLP ) and the effective dose ( ED ) in group B were lower than in group A 63.97%, 63.83% and 63.97%. The results of image quality evaluation: the mean CT value of group B was significantly higher than group A; The difference had statistical significance ( P〈0.05 ) . The difference of CNR, SNR hadn't statistical significant ( P〉0.05 ). And two groups' carotid reconstruction image quality, blood vessels, marginal form the overall satisfaction, no significant difference ( P〉0.05 ) .Conclusion: Using low tube voltage ( 80 kV ) can lower 64% radiation dose.It is proved to be effective and feasible in carotid CT angiography.
作者 金彤
出处 《中国医学创新》 CAS 2013年第28期79-82,共4页 Medical Innovation of China
关键词 体层摄影术 辐射剂量 管电压 图像质量 Computer tomography Radiation dose Tube voltage Image quality
  • 相关文献

参考文献10

  • 1葛全序,毕可森,丛培霞,王培玺,赵锦桥.80kV64排多层螺旋CT低辐射肺动脉成像[J].中华放射医学与防护杂志,2009,29(5):535-539. 被引量:44
  • 2袁庆海,刘建华,韩冰,杨哲,弓婷婷.应用80 kV管电压降低主动脉CT造影射线剂量初烫[J].中华放射医学与防护杂志,2010(6):740-743. 被引量:13
  • 3何怡红,罗天友.多层螺旋CT头颈部血管成像的应用进展[J].重庆医学,2007,36(9):884-887. 被引量:15
  • 4Menzel H,Schibilla H,Teunen D.Guidelines on radiation dose onthe patient[M].European Guidelines on Quality Criteria or ComputedTomography,2006:32-32.
  • 5Mettler F A,Bhargavan M,Faulkner K,et al.Radiologic and nuclearmedicine studies in the United States and worldwide:frequency,radiation dose,and comparison with other radiation sources-1950-2007[J].Radiology,2009,253(2):520-531.
  • 6Farr R F,Robcrts P J.Radiationphysics in Physics for medicalimagmg[M].London,England:Elsevier Science,2002:19-22.
  • 7Waaijer A,Prokop M,Velthuis B K,et al.Circle of Willis at CTangiography:dose reduction and image quality-reducing tube voltageand increasing tube current settings[J].Radiology,2007,242(3):832-839.
  • 8Szucs-Farkas Z,Kurmann L,Strautz T,et al.Patient exposureand image quality of low dose pulmonary computed tomographyangiography:comparison of 100 and 80 kVp protocols[J].Invest Radiol,2008,43:871-876.
  • 9Heyer C M,Mohr P S,Lemburg S P,et al.Image quality andradiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol[J].Prospective Randomized Study Radiology,2007,245(6):577-583.
  • 10Winter mark M,Maeder P,Verdun F R,et al.Using 80 kVp inperfusion CT measurement of regional cerebral blood flow[J].AJNRAm J Neuroradiol,2000,21(1):1881-1884.

二级参考文献53

共引文献63

同被引文献21

引证文献3

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部