期刊文献+

胸部肿瘤斜野照射治疗首次摆位射野参数不符的探讨

The exploration on the chest tumor oblique wild first position inconsistent treatment beam parameters
下载PDF
导出
摘要 目的:对胸部肿瘤斜野照射治疗首次摆位射野参数不符进行探讨,通过患者体表标记和升床标记提高摆位精度。方法:选择加速器治疗的40例胸部肿瘤斜野照射的患者首次摆位核对升床,照射野与模拟定位时的数据进行对比。结果:通过对患者首次摆位与模拟定位射野参数核准时升床高度误差为3~8mm;按医嘱升床及射野参数摆位,射野边界误差为3~9mm。结论:通过模拟定位射野参数摆位与体表标记不符的现象,寻找摆位误差的原因。 Objective: To explore the chest tumor oblique wild first position inconsistent treatment beam parameters, it can improve the positioning accuracy. Methods: The accelerator 40 thoracic tumors oblique field irradiation in patients with the first positioning liter bed radiation field with analog positioning data comparison. Results: Bit analog positioning patients for the first time placed a high degree of error of 3 to 8mm; liter bed when the beam parameters approved by your doctor liter bed and beam parameters anyway, the the beam boundary error in the 5 to 9mm. Conclusion: Positioning the beam parameters by simulating phenomena that are inconsistent with the surface markers positioning, looking for a positioning error.
出处 《中国医学装备》 2013年第10期106-107,共2页 China Medical Equipment
关键词 胸部肿瘤 等中心照射 模拟定位 摆位误差 Thoracic tumor Isocenter Irradiation Simulation Position error
  • 相关文献

参考文献5

二级参考文献34

  • 1申爱萍,周海,周晶,张翔.体部三维适形放射治疗中的摆位技术体会[J].肿瘤基础与临床,2006,19(1):65-66. 被引量:4
  • 2邵明海,胡炜,余文杰,于长辉,郑永敏.胸腹体架热塑体膜固定摆位误差分析[J].浙江临床医学,2006,8(4):433-434. 被引量:8
  • 3王义革,王忠建,陈平.激光定位系统在肿瘤放射治疗中的作用[J].医用放射技术杂志,2006(5):110-110. 被引量:3
  • 4沈君姝,王健琪,张宜勤,翟振宇,荆西京.头颈部肿瘤调强放疗中摆位偏差的测量与分析[J].第四军医大学学报,2006,27(12):1147-1149. 被引量:24
  • 5胡逸民,杨定宇.肿瘤放射治疗技师讲义[M].北京:中华医学会放射肿瘤治疗学学会,中国医学科学院肿瘤医院放射科,1993:130-144.
  • 6Hurkmans C W,Remeijer P,Lebesque JV,et al.Set-up verification using portal imagine; reive of currert clinical practice[J].Radiather Oncot,2005,58(2):105-120.
  • 7Samuelsson A,Mercke C,Johansson K A. Systematic set-up er rors for IMRT in the head and neck region: effect on dose distribution[J]. Radiother Oncol,2003, 66(3) :303-311.
  • 8Beckham W A, Keall P J, Siebers J V. A fluence convolution method to calculate radiation therapy dose distributions that incorporate random set-up error[J]. Phys Med Biol, 2002,47 (19):3465-3473.
  • 9Stroom J C,de Boer J C,Huizenga H,et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probility[J]. Int J Radiat Oncol Biol Phys, 1999,43(4):905-919.
  • 10Van Herk M,Remeijer P,Rasch C,et al. The probility of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy[J]. Int J Radiat Oncol Biol Phys,2000,47(4) :1121-1135.

共引文献37