Sysmex CA-7000全自动血凝分析系统检测项目PT、APTT、FIB性能验证实验分析 被引量:4

Verification experimental analysis on test performance of PT,APTT and FIB via Sysmex CA-7000 Automatic Blood Coagulation Analysis System
下载PDF
导出
摘要 目的验证Sysmex CA-7000全自动血凝分析系统对血浆凝血酶原时间(PT)、活化部分凝血活酶时间(APTT)、纤维蛋白原(FIB)的分析性能。方法按照临床和实验室标准化研究院(CLSI)的标准和指南文件对Sysmex CA-7000全自动血凝分析仪检测PT、APTT、FIB的精密度、正确度、分析测量范围、生物参考区间及携带污染率进行了验证,并与质量目标要求和厂商声明的分析性能进行比较。结果批内精密度<1/4允许总误差(Tea)(CLIA,88),批间精密度<1/3Tea(CLIA’88),正确度试验结果满足1/2Tea(CLIA’88);分析测量范围项目FIB结果相关性好(r>0.975),线性理想(斜率接近于1);生物参考区间及携带污染率与厂商提供的相符合。结论 Sysmex CA-7000全自动血凝分析仪检测项目PT、APTT、FIB在精密度、正确度、生物参考区间、分析测量范围、携带污染率等5个方面满足质量目标要求和厂家声明的分析性能。 Objective To verify the analytic performance of plasma prothrombin time ( PT), activated partial thromboplastin time (APTT) and fibrinogen (FIB) by Sysmex CA-7000 Automatic Blood Coagulation Analysis System. Methods According to the standards and the guidelines of the Clinical and Laboratory Standards Institute (CLSI) , we tried to verify the precision, trueness, analytical measurement range, biological reference interval and carry-over rates of PT, APTT and FIB, all of which were detected by Sysmex CA-7000 Automatic Blood Coagulation Analysis System. The results were compared with that of the quality target require-ments and the manufacturer statement analysis performance. Results The Intra-assay precisions were all less than 1/4 Tea ( CLIA' 88) , and the Inter-assay precisions were all less than 1/3 Tea ( CLIA' 88). Trueness test results met 1/2 Tea ( CLIA' 88 ). The analytical measurement range test in FIB was good ( r 〉 0. 975 ) , as well as the linearity ( slope was close to 1 ). The bio-logical reference interval and the carry-over rate were consistent with what the manufacturer provided. Conclusion The outcomes of PT, APTT and FIB detected by Sysmex CA-7000 Automatic Blood Coagulation Analysis System meets the quality target require-ments and the manufacturer statement analysis performance in all the five aspects including precision, trueness, analytical measure-ment range, biological reference interval and carry-over rate.
机构地区 解放军第
出处 《临床军医杂志》 CAS 2013年第10期1066-1068,共3页 Clinical Journal of Medical Officers
关键词 全自动血凝分析仪 性能评价 精密度 正确度 分析测量范围 生物参考区间 携带污染率 Automatic Blood Coagulation Analysis System performance verification precision trueness analytical measurementrange biological reference interval carry-over rate
  • 相关文献

参考文献8

  • 1Department of Health and Human Services. Centers for Medicare & Medicaid Services. Clinical laboratory improvement amend- ments of 1988 [S]. Fina Rule, Fed Register,2003:3704.
  • 2Clinical and laboratory Standards Institute. Evaluation of preci- sion performance of quantitative measurement methods [ S ]. EP15 - A2, CLSI,2004.
  • 3Clinical and laboratory Standards Institute. Evaluation of preci- sion performance of quantitative measurement methods [ S ]. EP9 - A2, CLSI,2002.
  • 4Clinical and laboratory Standards Institute. Evaluation of preci- sion performance of quantitative measurement methods[ S ]. EP6 - A2, CLSI,2003.
  • 5Clinical and laboratory Standards Institute. Evaluation of preci- sion performance of quantitative measurement methods [ S ]. C25 - A2, CLS1,2000.
  • 6Clinical and laboratory Standards Institute. Evaluation of preci- sion performance of quantitative measurement methods [ S ]. EP10 - A2, CLSI,2002.
  • 7王治国.临床检验方法确认与性能验证[M].北京:人民卫生出版社.2011:164.
  • 8杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.

共引文献173

同被引文献32

引证文献4

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部