320排CT对心房颤动患者冠状动脉CT成像的研究 被引量:4

The effectiveness of 320-detector CT coronary angiography in patients with atrial fibrillation
下载PDF
导出
摘要 目的:评价320排CT对心房颤动患者冠状动脉CT成像的有效性。方法:连续入组34例持续心房颤动患者。所有患者均进行了320排CTCA和常规冠状动脉造影(CCA)检查。两位评价者评价CTCA是否存在冠状动脉显著狭窄(≥50%),以CCA为参照标准,评价CTCA的敏感性、特异性、阳性预测值(PPV)和阴性预测值(NPV)。结果:以冠状动脉段为单位,CTCA的敏感性、特异性、PPV和NPV分别为94.1%、99.3%、84.2%和99.8%。结论:320排CTCA诊断的准确性较高,在心房颤动患者可以有效地排除冠心病。 Objective:To evaluate the effectiveness of CT coronary auography (CTCA) in patients with atrial fibrillation (AF) using 320-detector CT. Methods:Thirty-four patients with persistent AF were con- secutively enrolled in this study. All patients underwent 320 - detector CTCA and conventional coronary angiog- raphy (CCA). The presence of significant stenosis ( I〉50% ) were evaluated by two radiologists blinded to the results of CCA. The sensitivity, specificity, positive predictive value ( PPV), and negative predictive value (NPV) were calculated using CCA as the reference standard. Results: On per-segment analysis, sensitivity, specificity, PPV, and NPV were 94. 1%, 99. 3%, 84. 2%, and 99. 8%. Conclusion: 320-detector CTCA is accurate to exclude CAD in patients with AF.
出处 《心肺血管病杂志》 CAS 2013年第5期614-617,共4页 Journal of Cardiovascular and Pulmonary Diseases
基金 北京市优秀人才项目(2011D003034000030) 北京市卫生系统高层次卫生技术人才资助项目
关键词 体层摄影术 X线计算机 冠状动脉 心房颤动 Tomography X-ray computed Coronary artery Atrial fibrillation
  • 相关文献

参考文献1

共引文献128

同被引文献42

  • 1戴汝平,白桦,吕滨,张少雄,荆宝莲,曹程,何沙,任力.超高速CT在心血管病诊断中的应用[J].中华放射学杂志,1997,31(2):81-85. 被引量:52
  • 2吴文辉,吕滨,蒋世良,鲁锦国,乔树宾,吴永健,戴汝平,沈云.多层螺旋CT对冠状动脉粥样硬化斑块的显示结果及与超声结果的比较[J].中华放射学杂志,2007,41(10):1027-1031. 被引量:24
  • 3Moscariello A, Takx RA, Schoepf U J, et al. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection. Eur Radiol, 2011, 21(10): 2130-2138.
  • 4Dewey M, Zimmermann E, Deissenrieder F, et al. Noninvasive coronary angiography by 320-row computed tomography with lower radiation exposure and maintained diagnostic accuracy: comparison of results with cardiac catheterization in a head-to- head pilot investigation. Circulation, 2009, 120(10): 867-875.
  • 5Defrance T, Dubois E, Gebow D, et al. Helical prospective ECG-gating in cardiac computed tomography: radiation dose and image quality. Int J Cardiovasc Imaging, 2010, 26(1): 99- 107.
  • 6Mulkens TH, Bellinck P, Baeyaert M, et al. Use of an automatic exposure control mechanism for dose optimization in multi- detector row CT examinations: clinical evaluation. Radiology, 2005, 237(1): 213-223.
  • 7Nishimaru E, Ichikawa K, Hara T, et al. Novel method of noise power spectrum measurement for computed tomography images with adaptive iterative reconstruction method. Nihon Hoshasen Gijutsu Gakkai Zasshi, 2012, 68(12): 1637-1643.
  • 8Yoo RE, Park EA, Lee W, et al. Image quality of adaptive iterative dose reduction 3D of coronary CT angiography of 640-slice CT: comparison with filtered back-projection. Int J Cardiovasc Imaging, 2013, 29(3): 669-676.
  • 9Austen WG, Edwards JE, Frye RL, et al. A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation, 1975, 51 (4 Suppl): 5-40.
  • 10European Commisson. European guidelines on quality criteria for computed tomography. Report EUR 16262 EN. http://www. drs.dk/guidelines/ct/quality/index.html.

引证文献4

二级引证文献18

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部