期刊文献+

量子收率测定仪

An Instrument for Photochemical Reaction Study
下载PDF
导出
摘要 本文报导用于光化学反应研究的量子收率测定仪.该仪器主要由200瓦氙灯光源和带恒温夹套的石英反应器等十几个部件构成.核心部件光反应器的两个窗口和反应器壁烧结成为一个整体,可耐强酸、强碱和各种溶剂.采用草酸铁钾化学光度计测量单色光强度,可达2~8×10^(-8)爱因斯坦/秒,相对误差为±3%;测量催化光还原水产氢体系的量子收率,相对误差为±10%,在可比条件下与文献结果一致. An instrument for photochemical reaction study, is reported. The instrument consists of a 200W xenon light source, a 40mL quartz photoreactor and several other parts. Two quartz windows are sintered with the walls of the photoreactor to form an integrated part. The intensity of the monochromatic light, generally 2~8×10^(-8) Einstein/s, is measured by a ferrioxalate chemical actinometer. The maximum deviations are within±3%.The relative errors for determination of ΦH_2 of two catalytic photoreduction watergenerated hydrogen systems are ±10%.
出处 《分析仪器》 CAS 1991年第4期34-37,共4页 Analytical Instrumentation
  • 相关文献

参考文献4

  • 1史晓波,赵成文,顾长立.不同条件下入射光波长对产氢量子收率的影响[J]太阳能学报,1988(01).
  • 2史晓波,赵成文,汤焕毅,顾婉贞,顾长立.量子收率的影响[J]太阳能学报,1986(01).
  • 3赵成文,史晓波,汤焕毅,顾婉贞,顾长立.若干因素对催化光还原水产氢的量子收率的影响[J]催化学报,1985(01).
  • 4赵成文,史晓波,戚奎华,汤焕毅,顾婉贞.几种光还原水催化体系产氢量子收率的研究[J].太阳能学报,1983,9(1):25-29. 被引量:1