口腔不同金属材料全冠修复体CT成像伪影的实验研究 被引量:2

CT artifact caused by various types of dental crown in oral cavity
下载PDF
导出
摘要 目的:对口腔常用的金属材料全冠在螺旋CT图像中的伪影进行实验研究,为临床口腔金属的选择提供参考依据。方法:用同一家猪头颅(六月龄,洛克白猪),模拟人的在体口腔环境,使用金钯、银钯、纯钛、钴铬、镍铬5种金属材料,在同一牙位制作5种金属冠,并分别戴入头颅的右下第1、2、3前磨牙,按照常规颌骨CT条件扫描(130ekv,1mm),测量伪影的尺寸、形态和面积。结果:5种金属冠都会产生以冠为中心向四周散射的伪影,金属的密度越高高亮度部分伪影面积越大,沿投照方向的长度越大,与原始截面图差异越大,差异具有统计学意义(P<0.05)。不同金属所产生的高亮度伪影和条索状伪影排列顺寻为纯钛<镍铬<钴铬<金钯<银钯,差异具有统计学意义(P<0.05)。结论:冠修复材料应首选密度较低的金属。 Objective: To compare various types of dental crown in CT images in order to provide evidence for choosing metallic dental crowns. Methods. Five kinds of metal materials widely used as dental prostheses including gold, silver, pure titanium, cobalt-chromium, nickel-chromium were used as dental crowns worn into the 1st, 2nd and 3rd right premolars in the cavity of the same pig (six months, Locke white pig). Conventional jaw CT scans were applied (130 ekv, 1 mm) to measure the size, shape and vol- ume of crown artifacts. Results: All the five types of metal crowns produced scattering artifacts surround- ing the crown center, and the brightness, ing significant difference comparing with brightness and length of the artifacts, the cobalt-chromium 〈 gold 〈 silver (P 〈 0.05 )crown repair material. length of the artifacts increased with the density of metals show- the original CT image (P%0.01). According to the order of five metals can be listed as. pure titanium %nickel-chromium 05). Oonclusion: metal of low density should be preferred for
出处 《海南医学院学报》 CAS 2013年第12期1764-1766,1770,共4页 Journal of Hainan Medical University
基金 中国高校医学期刊临床专项资金项目(112210912)~~
关键词 计算机断层扫描 金属材料 伪影 Computed tomography scan, Metallic materials, artifacts
  • 相关文献

参考文献5

二级参考文献70

  • 1高鸿奕,谢红兰,陈建文,李儒新,徐至展,田玉莲,朱佩平,冼鼎昌.硬X射线相位衬度成像的实验研究[J].中国激光,2005,32(2):167-169. 被引量:14
  • 2方正,孙小敏,骆清铭.一种简便的计算层析系统X射线硬化校正方法[J].光学学报,2007,27(2):302-306. 被引量:4
  • 3Dalai T, Kalra MK, Rizzo SM, et al. Metallic prosthesis: technique to avoid increase in CT radiation dose with automatic tube current modulation in a phantom and patients. Radiology. 2005:236(2):671-675.
  • 4White LM, Kim JK, Mehta M, et al. Complications of total hip arthroplasty: MR imaging-initial experience. Radiology. 2000; 215(1 ):254-262.
  • 5Fishman EK, Magid D, Robertson DD, et al. Metallic hip implants: CT with multiplanar reconstruction. Radiology. 1986; 160(3): 675-681.
  • 6Vande Berg B, Malghem J, Maldague B, et al. Multi-detector CT imaging in the postoperative orthopedic patient with metal hardware. Eur J Radiol. 2006;60(3):470-479.
  • 7Kovacs M, Fejerdy P, Dobo NC. Metal artefact on head and neck cone-beam CT images. Fogorv Sz. 2008;101 (5): 171-178.
  • 8d'Agostino AG, Remy-Jardin M, Khalil C, et al. Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients. Eur Radiol. 2006; 16(10):2137-2146.
  • 9Kim JC, Kim JH, Kim HC, et al. Effect of fiber structure on dialysate flow profile and hollow-fiber hemodialyzer reliability: CT perfusion study. Int J Artif Organs. 2008;31 (11):944-950.
  • 10Lim DS, Lee IS, Choi K J, et al. The potential for non-invasive study of mummies: validation of the use of computerized tomography by post factum dissection and histological examination of a 17th century female Korean mummy. J Anat. 2008;213(4):482-495.

共引文献30

同被引文献19

引证文献2

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部