目的消除右胸电阻抗容积波中呼吸的干扰。方法首先讨论了对传统抗呼吸干扰的相干平均法改进 ,同时提出了估计拟合呼吸波和时域切比雪夫正交多项式消除呼吸干扰的方法。结果 3种处理方法结果表明 ,估计拟合呼吸波方法在实时性和保持逐拍信息性方面要优于其他两种。结论解决了实时监护的难点。
Objective To find a method of reducing Respiratory Interference in measurement of right chest impedance. Method The improvement of coherent averaging method of Reducing Respiratory Interference was discussed,and evaluation-fitting respiratory wave and filter of Chebyshev's polynomial as presented. Result The result indicated that evaluation-fitting of respiratory wave is better than other methods in its character of real time and keeping pulse information. Conclusion It solved difficulty in real time monitoring.