摘要 目的 利用自合成新色原测定血清葡萄糖.方法 采用双试剂速率法.结果 方法重复性:批内CV=0.84%,批间CV=1.2%,回收率99.7%,方法线性可达65mmol/L.本法能有效排除标本溶血、黄疸、内源性还原性物质及脂浊等干扰物对血清葡萄糖测定的影响,结果与GOD-POD终点法和已糖激酶速率法有良好相关,相关系数分别为0.996和0.998.结论 本方法准确可靠,重复性好,试剂稳定,线性较宽,值得应用.  Objective Using self building-up new chromogen for determination of glucose in serum. Methods Used double-reagent and kinetics. Results The within-run CV and between-run CV were 0.84% and 1.2% respectively. The recovery rate of glucose was 99.7% linearity up to 65mmol/L. Adverse factors interfering with determination of serum glucose such as hemolysis bilirubin endogenous desoxidative substance and lipid-micelle in serum were effectively eliminated. The results of the method correlated well with those of GOD-POD and glucose kinase methods. The correlation coefficent were 0.996 and 0.998 respectively. Conclusion This method is accurate credible stable repeatable and can be used in automatic biochemical analyzer with double-reagents. 
出处 《江西医学检验》 2000年第4期206-208,共3页 Jiangxi Journal of Medical Laboratory Sciences
关键词 葡萄糖 速率法 新色原双试剂 血清诊断  Glucose Kinetics Double-reagent New chromogen