Myoton-3肌肉检测仪在健康成人肌张力测量中的评价者间信度 被引量:15

Inter-rater Reliability of Myoton-3 Myometer for Assessing Muscle Tone in Healthy Adults
下载PDF
导出
摘要 目的:检验Myoton-3肌肉检测仪用于健康成人肌张力定量测量的评价者间信度。方法两名经过培训的测评人员分别使用Myoton-3对20名健康志愿者双侧肱二头肌、桡侧腕屈肌肌张力在24 h内进行测试:采用三次扫描和十次扫描模式,测得的肌肉放松状态下阻尼振动频率(F值)。计算组内相关系数(ICC),进行Bland-Altman分析。结果两名测试者间三次扫描模式所得F值的ICC为0.72~0.88,十次扫描模式测得F值的ICC为0.79~0.89。Bland-Altman分析显示,测试者间的一致性满足临床需要。结论 Myoton-3在健康成人上肢肌张力测量中具有良好的评价者间信度。 Objective To examine the inter-rater reliablity of Myoton-3 Myometer in assessment of muscle tone in healthy subjects. Methods 20 healthy volunteers were assessed their muscle tone of biceps brachii and flexor carpi radialis muscles in relaxed state with Myo-ton-3 by 2 testers within 24 h. The frequency of damping oscillations (F value) measured by Myoton-3 was as the characteristic of the mus-cle tone. The intraclass correlation coefficient (ICC) and Bland-Altman analysis were performed. Results The ICC was 0.72~0.88 and 0.79~0.89 in triple scan and ten-time scan pattern, respectively. The Bland-Altman analysis revealed no systematic errors between testers. Conclu-sion The Myoton-3 Myometer is reliable between testers for measuring the muscle tone in healthy adults.
出处 《中国康复理论与实践》 CSCD 北大核心 2013年第11期1058-1060,共3页 Chinese Journal of Rehabilitation Theory and Practice
基金 中山大学医科2009年第一批学生业余科研暨科研助理计划项目
关键词 Myoton-3 肌张力 评价者间信度 Myoton-3 muscle tone inter-rater reliablity
  • 相关文献

参考文献14

  • 1燕铁斌,窦祖林,冉春风.实用瘫痪康复[M].2版.北京:人民卫生出版社,2010:124-125.
  • 2Mutlu A, Livanelioglu A, Gtmel MK. Reliability of Ashworth and Modified Ashworth Scales in children with spastic cerebral palsy [J]. BMC Muskuloskelet Disord, 2008, 9: 44.
  • 3纪树荣,杨今姝.等速运动测试仪量化评定痉挛的研究[J].中国康复理论与实践,1999,5(2):75-79. 被引量:6
  • 4Bizzini M, Mannion AF. Reliability of a new, hand-held device for assessing skeletal muscle stiffness [J]. Clin Biomech, 2003, 18(5): 459-461.
  • 5Veldi M, Vasar V, Vain A, et al. Obstructive sleep apnea and ageing: Myotonometry demonstrates changes in the soft palate and tongue while awake [J]. Pathophysiology, 2004, 11(3): 159-165.
  • 6Viir R, Laiho K, Kramarenko J, et al. Repeatability of trapezius muscle tone assessment by a myometric method [J]. J Mecha- nisms Med Biol, 2006, 6(2): 215-228.
  • 7李镒冲,李晓松.两种测量方法定量测量结果的一致性评价[J].现代预防医学,2007,34(17):3263-3266. 被引量:153
  • 8潘晓平,倪宗瓒.组内相关系数在信度评价中的应用[J].华西医科大学学报,1999,30(1):62-63. 被引量:103
  • 9陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计,2007,24(3):308-309. 被引量:216
  • 10Roja Z, Kalkis V. Assessment of skeletal muscle fatigue of road maintenance workers based on heart rate monitoring and myotonometry [J]. J Occup Med Toxicol, 2006, 1 : 20.

二级参考文献22

  • 1赵希刚,李坤成,李永忠.单次屏气电影法MRI评价左心室功能的初步研究[J].中国医学影像技术,2005,21(1):65-67. 被引量:12
  • 2陈玉平,刘雪琴,蔡德鸿.骨质疏松症知识问卷的信度和效度测定[J].中国骨质疏松杂志,2005,11(3):339-341. 被引量:140
  • 3Bailey KD 许真(译).现代社会研究方法[M].上海:上海人民出版社,1986.90-106.
  • 4潘宝骏,科研通用电脑软件,1995年,259页
  • 5柯惠新,调查研究中的统计分析法,1992年,357页
  • 6苏驼,社会调查研究方法,1989年,66页
  • 7许真(译),现代社会研究方法,1986年,90页
  • 8Bland JM,Altman DG.Statistical methods for assessing agreement between two methods of clinical measurement.Lancet,1986,i:307-310.
  • 9Bland JM,Altman DG.Measuring agreement in method comparison studies.Statis tical Methodsin Medical Research,1999,8:135-160.
  • 10Najib Aziz,Michael RI,Sally SD,et al.Interpreting Method Comparison Studiesby Useofthe Bland-Altman Plot:Reflecting the Importance of Sample Sizeby Incorporating Confidence Limitsand Predefined Error Limitsinthe Graphic.Clinical Chemistry,2004,50 (11):2216-2222.

共引文献423

同被引文献132

引证文献15

二级引证文献92

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部