摘要
目的建立RP—HPLC—uV法用于盐酸雷诺嗪的含量和有关物质测定。方法采用WatersXBRIDGEC。8色谱柱(250mm×4.6mm,5μm),流动相为乙腈一20mmol·L^-1磷酸二氢钾溶液(用三乙胺调节pH值至7.0)(体积比40:60),流速为1.0mL·min^-1,检测波长为215nm,进样量为10μL。结果盐酸雷诺嗪质量浓度在9.068—226.7mg·L^-1内线性关系良好(r=0.9996),重复性试验RSD为0.96%,主峰与各杂质均具有较好的分离度。结论该方法快速、准确、专属性强,适用于盐酸雷诺嗪及其制剂的质量控制。
Objective To establish an HPLC-UV method for the detection of ranolzaine hydrochloride and re- lated compounds. Methods A Waters XBRIDGE C18 column (250 mm ×4.6mm,5μm) was used;the mo- bile phase consisted of acetonitrile-KH2PO4 solution (20 mmol. L ^-1 , adjusted to pH 7.0 with triethylamine) ( V: V=40:60) ;the flow rate was set at 1.0 mL.min^-1 and the detector was set at 215 nm with an injection volume of 10 μL. Results The linear range was 9. 068-226. 7 mg.L^-1 (r =0. 999 6). The relative standard deviation for precision was 0. 2%. The resolution between ranolazine and corresponding related substances was good. Conclusions The present method is rapid, accurate, sensitive, and suitable for the quality control of ranolazine hydrochloride and its formulations.
出处
《沈阳药科大学学报》
CAS
CSCD
北大核心
2013年第11期872-875,共4页
Journal of Shenyang Pharmaceutical University
关键词
盐酸雷诺嗪
含量测定
有关物质
高效液相色谱
ranolzaine hydrochloride
content determination
related substance
HPLC