摘要
建立巴鱼中呋喃唑酮、呋喃它酮、呋喃西林和呋喃妥因的代谢物残留量液相色谱-串联质谱的方法。分析物采用电喷雾电离正离子( ESI +)、多反应监测(M RM )模式检测,内标法定量。本实验对固相萃取法提取和乙酸乙酯提取净化前处理方法进行比对,4种硝基呋喃代谢物的混合标准溶液的浓度在0.5~50ng/mL范围线性良好,相关系数均大于0.9981,4种化合物的检测限为0.5μg/kg。添加浓度1μg/kg、2.5μg/kg、5μg/kg范围内得出4种代谢物固相萃取净化法的回收率在81.1%~104.1%之间,相对标准偏差(RSD)小于5.7%.乙酸乙酯提取净化法的回收率在88.5%~108.2%之间,相对标准偏差(RSD )小于6.1%。结论证明乙酸乙酯提取法方法前处理简单、快速、稳定性好,适用于大量样品的同时检测。
A method was established to detect nitrofuran metabolic residues in Spanish mackerel by liq-uid chromatography tandem mass spetrometry (LC-MS/MS) .Theanalvtes were detected by LC-MS/MS (ESI) with multiple reaction monitoring (MRM ).The result indicated that the calibration curves were in good linearity between the peak area and the concentration of 0.5-50ng/ml for nitrofuran metabolic with correlation cofficients not less than 0.9981.The limits of detection for four nitrofuran metabolics were 0.5μg/kg. .Then analytes were extracted by spe or ethyl acetat the average recoveries were 81.1%-104.1% or 88.5%-108.2% and the related standard deviations less than 5.1 or 6.1% at the three levels 1μg/kg、2.5μg/kg、5μg/kg.The proposed method was simple ,rapid ,the stability was good and suitable for the simultaneous quantitative detection.
出处
《分析仪器》
CAS
2013年第6期92-97,共6页
Analytical Instrumentation
基金
中山市科技计划项目(20114A192)