期刊文献+

模拟体外循环下肝素涂层的稳定性实验

Study of heparin stability on heparin-coated circuits under simulating cardiopulmonary bypass
下载PDF
导出
摘要 目的:评估肝素涂层体外循环管道肝素的稳定性。方法:测定肝素图层体外循环管道模拟体外循环法与浸泡法在不同时点的肝素脱落率。结果:在128 h内,两组肝素结合表面均有不同程度的肝素分子脱落现象,肝素脱落率随时间增大,在一定时间内浸泡法结合的肝素表面的肝素最大脱落率为85%,而体外循环法结合表面的肝素脱落率为92%。体外循环组涂层表面中肝素的脱落率与浸泡组两者之间无显著差异。结论:短期内肝素结合体外循环管道比较牢固,满足临床使用。 AIM: To evaluate heparin stability on heparin-coated cardiopulmonary bypass (CPB) circuits. METHODS: Heparin off-rates of CPB stimulation method and immersion method at different time points were determined. RESULTS: At 128 h, different levels of heparin molecule shedding phenomenon were observed in heparin-binding surface in both groups and the heparin off-rate increased with time. The maximum heparin off-rates of the CPB method and immersion method were, respectively, 92 and 85 during a certain period of time, with no significant difference between groups. CONCLUSION: Heparin stability on heparin-coated circuits of CPB has relatively stronger heparin binding and is of better clinical use.
出处 《心脏杂志》 CAS 2013年第5期573-575,共3页 Chinese Heart Journal
基金 国家自然科学基金项目资助(30700174) 陕西省社发攻关基金[(2006K12-G5(4)]项目资助
关键词 肝素涂层管道 脱落率测定 肝素稳定性 heparin-coated circuits heparin off-rates heparin stability
  • 相关文献

参考文献6

二级参考文献41

  • 1张安兄,吕德龙,钟伟,程为庄,杜强国.生物材料的血液相容性[J].上海生物医学工程,2004,25(3):53-58. 被引量:27
  • 2杨剑,易定华,刘金成,刘剑,曹瑞军,秦锋.肝素涂层医用聚氯乙烯的细胞毒性研究[J].生物医学工程学杂志,2005,22(4):739-741. 被引量:5
  • 3[1]Sperling C,Houska M,Brynda E,et al.In vitro hemocompatibility of albumin-heparin multilayer coatings on polyethersulfone prepared by the layer-by-layer technique[J].J Biomed Mater Res A,2006,76(4):681-689.
  • 4[2]Wendel HP,Ziemer G.Coating-techniques to improve the hemocompatibility of artificial devices used for extracorporeal circulation[J].Eur J Cardiothorac Surg,1999,16(3):342-350.
  • 5[4]Hsu LC.Biocompatibility in eardiopulmonary bypass[J].J Cardicthoracic Vasc Anesth,1997,11(3):376-382.
  • 6[7]Huang LY,Yang MC.Hemocompatibility of layer-by-layer hyaluronic acid/heparin nanostructure coating on stainless steel for cardiovascular stents and its use for drug delivery[J].J Nanosci Nanotechnol,2006,6(9-10):3163-3170.
  • 7[8]Thorslund S,Sanchez J,Larsson R,et al.Functionality and stability of heparin immobilized onto poly(dimethylsiloxane)[J].Colloids Surf B Biointerfaces,2005,45(2):76-81.
  • 8[9]Lappegard KT,Riesenfeld J,Brekke OL,et al.Differential effect of heparin coating and complement inhibition on artificial surface-induced eicosanoid production[J].Ann Thorac Surg,2005,79(3):917-923.
  • 9Sperling C, Houska M, Brynda E, et al. In vitro hemocompalibility of albumin-heparin muhilayer coatings on polyethersulfone prepared by the layer-by-layer technique[ J ]. J Biomed Mater Res A, 2006, 76(4) : 681 -689.
  • 10Wendel HP, Ziemer G. Coating-techniques to improve the hemocompatibility of artificial devices used for extracorporeal circulation [J]. Eur J Cardiothorac Surg, 1999, 16(3) : 342 -350.

共引文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部