三维剂量验证系统Delta4在容积旋转调强计划剂量验证中的应用 被引量:6

Application of 3D Dosimetric verification system Delta4 for VMAT verification
下载PDF
导出
摘要 目的:探讨三维剂量验证系统Delta4对容积旋转调强计划进行剂量验证的可行性。方法:利用Delta4对20例患者的容积旋转调强计划进行剂量验证;为了对比,同时用MatriXX电离室矩阵对相同计划的冠状面进行剂量分布验证;利用剂量仪、指形电离室对相同计划进行点剂量的绝对剂量验证;将三种方法的验证结果进行比对。剂量分布验证采用Gamma分析,取剂量3%、距离3mm,通过率≥95%为验证通过标准。结果:三种验证方法所有计划的误差均在允许范围之内。使用Delta4验证治疗计划的Gamma通过率平均为98.90%,各个射野的Gamma通过率平均99.83%;用MatriXX电离室矩阵验证各计划冠状面剂量分布Gamma通过率平均为98.02%;绝对剂量验证平均误差为1.18%。结论:利用Delta4对旋转容积调强计划进行剂量验证是可行的。 Objective: To discuss the application of 3D dosimetric verification system Delta4 for VMAT verification. Methods: Evaluation the dosimelrie verification of VAMT plans of 20 patients with Delta4; For comparison, the Coronal plane does of the same plan were measured with MatriXX ionization chamber array; the absolute does of the same plan were measured with dosimeter and ionization chamber. Compare the results of the three methods. The results of the dose distribution were analyzed by Gamma analysis, with dose 3%, distance 3ram, the criterion of a successful verification is that the passing ratio reaches greater than 95%. Results: The errors of the three methods all within the allowance range. The overall average percentages of passing ratio measured with Delt4 for the plan and each field is 98.90% and 99.83%, respectively; The overall average percentages passing ratio measured with MatriXX and the average errors of the absolute does is 98.02% and 1.18%,respectively. Conclusions: The application of 3D dosimetric verification system Delta4 for VMAT verification is practicable.
出处 《中国医学物理学杂志》 CSCD 2013年第6期4497-4499,共3页 Chinese Journal of Medical Physics
关键词 容积旋转调强 Delta4 剂量验证 VMAT delta4 dosimetric verification
  • 相关文献

参考文献2

二级参考文献26

  • 1Otto K. Volumetric modulated arc therapy:IMRT in a single gantry arc[J].Medical Physics,2008,(01):310-317.doi:10.1118/1.2818738.
  • 2Bzdusek K,Friberger H,Eriksson K. Development and evaluation of an efficient approach to volumetric arc therapy planning[J].Mad Phys,2009,(06):2328-2339.
  • 3Bhagwat MS,Han ZH,Ng SK. An oscillating sweeping gap test for VMAT quality assurance[J].Phya Med Biol,2010,(17):5029-5044.
  • 4Alvarez-Moret J,Pohl F,Koelbl O. Evaluation of volumetric medulated arc therapy (VMAT) with Oncentra MasterPlan(⑧) for the treatment of head and neck cancer[J].Radiation Oncology,2010,(05):110-120.
  • 5Ran M,Yang WS,Chen F. Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT:plan quality,delivery efficiency and accuracy[J].Medical Physics,2010,(03):1350-1359.
  • 6Bortfeld T,Webb S. Single-arc IMRT[J].Physics in Medicine and Biology,2009,(01):N9-N20.
  • 7Yon G,Lu B,Kozelka J. Calibration of a novel four-dimensional diode array[J].Medical Physics,2010,(01):108-115.doi:10.1118/1.3266769.
  • 8Létourneau D,Publieover J,Kozelka J. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy[J].Medical Physics,2009,(05):1813-1821.
  • 9Yoda K,Nakagawa K,Shiraishi K. Dose verification of intensity modulated arc therapy using an ERGO + + treatment planning system and Elekta internal multileaf collimators for prostate cancer treatment[J].British Journal of Radiology,2009,(976):328-331.doi:10.1259/bjr/13535955.
  • 10Dobler B,Streck N,Klein E. Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber army I'mRT MatriXX—a feasibilily study[J].Physics in Medicine and Biology,2010,(02):N39-N55.

共引文献30

同被引文献36

引证文献6

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部