利用二维电离室矩阵验证食管癌调强放疗的剂量分布

Verification of dose distribution in intensive modulation radiotherapy( IMRT) for esophageal cancer patients with 2D ionization chamber array
下载PDF
导出
摘要 目的比较实际受照剂量分布与计划剂量分布,评价食管癌调强放射治疗计划处方实施的准确性。方法选取135例食管癌患者,采用束流调强放射治疗,并将其中99例治疗计划移植到二维电离室矩阵和验证模体中,生成验证计划,在加速器上进行模拟治疗。利用二维电离室矩阵进行测量,将测得剂量数据和计划剂量数据进行比较和分析。结果经过Gamma分析(按3%/3 mm的误差标准),82.8%的计划验证通过率≥90.0%,平均通过率为94.8%,标准误差为0.6%。结论实际测量剂量分布与计划剂量分布符合情况基本理想,部分治疗计划需进一步调试修正直至验证通过后才可用于临床治疗。 Objective To evaluate the accuracy of radiation dose plan in intensive modulation radiotherapy (IMRT) for patients with esophageal cancer. Methods One hundred and thirty five patients with esophageal cancer underwent IMRT. The close plans were designed and 99 dose plans were moved into 2D-An'ay ionization chamber and phantom. Assurance plans were developed and treatment was simulated on linear accelerator. Radiation doses were measured with 2D-Array ionization chamber and compared to the plans. Results Gamma analysis at the 3%/3 mm criteria revealed that the pass rates of 82.8% plans were ≥90.0% with an average rate of 94.8% ( SE : 0.6% ). Conclusion The consistency of measured dose distribution with the planned dose distribution is satisfactory, but some treatment plans need to be tested and modified before clinical application.
出处 《实用肿瘤杂志》 CAS 2013年第6期632-636,共5页 Journal of Practical Oncology
关键词 食管肿瘤 放射疗法 放射治疗剂量 矩阵 esophageal neoplasms/radiotherapy radiotherapy dosage MATRIX
  • 相关文献

参考文献5

  • 1胡逸民.调强适形放射治疗[A]北京:原子能出版社,1999538-541.
  • 2陆军,胡彩容,蔡勇君,阴晓娟,陈俊强.三维剂量验证系统在容积旋转调强中的应用[J].中华放射医学与防护杂志,2013,33(1):58-61. 被引量:4
  • 3Stock M,Kroupa B,Georg D. Interpretation and evaluation of the Gamma index and Gamma index angle for the verification of IMRT hybrid plans[J].Physics in Medicine and Biology,2005,(03):399-411.doi:10.1088/0031-9155/50/3/001.
  • 4Chao KSC;Ozhigit G;冯平柏.实用肿瘤调强放射治疗[M]南京:江苏科学技术出版社,200610-17.
  • 5张书旭,余辉,王诗琴,谭剑明,林生趣.MatriXX系统在放疗质控中的应用研究[J].中国辐射卫生,2009,18(3):272-274. 被引量:11

二级参考文献32

  • 1张书旭,徐海荣,陈光杰,杨克柽,林生趣,余辉.呼吸运动对靶区三维重建的影响[J].中国医学物理学杂志,2006,23(6):400-404. 被引量:14
  • 2Andenna C, Benassi M, Caccia B, et al. Comparison of dose distributions in IMRT planning using the gamma function [ J ]. J Exp Clin Cancer Res,2006,25 (2) :229 - 234.
  • 3Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel - segmented ionization chamber [J]. MedPhys,2004,31(2):414-420.
  • 4Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine[J]. Phys Med Biol,2007,52(4) : 1 197 - 1 208.
  • 5Jursinic PA, Nelms BE. A 2 - D diode array and analysis software for verification of intensity modulated radiation therapy delivery [ J ]. Med Phys, 2003,30 ( 5 ) : 870 - 879.
  • 6Wiezorek T, Banz N, Schwedas M. Dosimetric quality assurance for intensity - modulated radiotherapy feasibility study for a filmless approach [ J ]. Strahlenther Onkol, 2005, 181 (7) :468 -474.
  • 7Verbakel WF, Cuijpers JP, Hoffmans D, et al. Volumetric intensity-modulated arc therapy vs conventional IMRT in head- and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys,2009, 74( 1 ) :252-259.
  • 8Wolff D, Stieler F, Welzel G, et al. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-eonformal RT for treatment of prostate cancer. Radiother Oneol,2009, 93 ( 2 ) :226-233.
  • 9Stathakis S. Ionization chamber array for patient specific VMAT, Tomotherapy and IMRT QA. J Phys: Conf Ser, 2010, 250:012029.
  • 10Stasi M, Baiotto B, Barboni G, et al. The behavior of several rnicroionization chambers in small intensity modulated radiotherapy fields. Med Phys, 2004, 31 (10) :2792-2795.

共引文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部