摘要 提出一种远程糖尿病性视网膜病变(DR)筛查系统的设计方案,重点介绍了通信子系统。由ARM7微控制器与继电器等组成的串口通信模块完成处于远程终端的计算机与眼底照相机之间的连接;使用C++语言完成基于MFC的通信程序设计,程序包括诊疗与诊断信息模块、视音频监控模块及眼底照相机操控模块三个部分。本方案在与本系统相似的远程医疗解决方案中具有较强的通用性。 A design scheme of a tele-screening system for diabetic retinopathy (DR) has been proposed, especially the communication subsystem. The scheme uses serial communication module consisting of ARM 7 microcontroller and relays to connect remote computer and fundus camera, and also uses C+ + programming language based on MFC to design the communication software consisting of therapy and diagnostic information module, video/audio surveillance module and fundus camera control module. The scheme possesses universal property in some remote medical treat- ment systems which are similar to the system.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2013年第6期1141-1146,共6页 Journal of Biomedical Engineering
基金 国家自然科学基金(科学仪器基础研究专项基金)资助项目(60827002) 福建省教育厅教育科研项目资助(JB13148)
关键词 远程医疗 糖尿病性视网膜病变 眼底照相机 串口通信 视音频通信 Remote medical treatment Diabetic retinopathy (DR) Fundus camera Serial communication Video/audio communication
  • 相关文献

参考文献9

二级参考文献53

共引文献40

同被引文献21

  • 1ALSABTI K, RAIZADA S, WANI V B, et al. Efficacyand reliability of fundus digital camera as a screening toolfor diabetic retinopathy in Kuwait[ J]. Journal of Diabe-tes and its Complications, 2003 , 17(4) ; 229-233.
  • 2LEESE G P, ELLINGFORD A, MORRIS A D, et al.Screening using compressed digital retinal images suc-cessfully identifies retinopathy [ J ]. Diabetes Care,2003,26(1) : 247.
  • 3GARG S,DAVIS R M. Diabetic retinopathy screeningupdate[ J]. Clinical Diabetes, 2009 , 27(4) : 140-145.
  • 4MASSINP,CHABOUIS A,ERGINAY A, et al.OPHDIAT : A telemedical network screening system fordiabetic retinopathy in the lle-de-France[ J] . Diabetes &metabolism, 2008, 34(3) : 227-234.
  • 5SCHULZE-DOBOLD C, ERGINAY A, ROBERT N, etal. OPHDIAT ; Five-year experience of a telemedicalscreening programme for diabetic retinopathy in Paris andthe surrounding area[ J]. Diabetes & Metabolism, 2012,38(5) : 450457.
  • 6ZEIMER R,ZOU S,MEEDER T, et al. A fundus cam-era dedicated to the screening of diabetic retinopathy inthe primary-care physician ’ s office [ J ]. InvestigativeOphthalmology & Visual Science, 2002, 43 ( 5 ):1581-1587.
  • 7DAUGMAN J G. High confidence visual recognition ofpersons by a test of statistical independence [ J ]. IEEETransactions on Pattern Analysis and Machine Intelli-gence, 1993,15(11) : 1148-1161.
  • 8WILDES R P. Iris recognition: An emerging biometric technology [ J ]. Proceedings of the IEEE, 1997, 85 (9) : 1348-1363.
  • 9ZHU D,MOORE S T, RAPHAN T. Robust pupil centerdetection using a curvature algorithm [ J ]. ComputerMethods and Programs in Biomedicine, 1999 , 59 (3):145-157.
  • 10SHINODA T, KATO M. A pupil diameter measurementsystem for accident prevention [ C ]. IEEE InternationalConference on Systems, Man, and Cybernetics, Taipei,2006: 1699-1703.

引证文献1

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部