摘要 目的:对便携式血糖仪的性能指标进行质量控制检测,判断其测试数值是否可被接受。方法:利用NCCLS的评价体系为检测指导,利用全自动生化分析仪、己糖激酶法测得数据为对照,对血糖仪测得结果进行分析。结果:经过对便携式血糖仪进行质量控制检测分析,其精密度、准确性、抗干扰性均符合相关要求,表明该种方法可以被接受。结论:经过质量控制检测,便携式血糖仪可以较好地用于血糖测试。 Objective To perform quality control detection of the performance indexes of the portable blood glucose meter in order to determine if the measurements are acceptable. Methods NCCLS evaluation system was used for direction, and then the results by the meter were compared with those by fully automatic biochemical analyzer and hexokinase method. Results The precision, accuracy and anti interference performance of the meter could meet the requirements, and thus the blood glucose by the meter was feasible. Conolusion Quality control detection proves that the portable meter can be used for blood glucose measuring.
机构地区 解放军
出处 《医疗卫生装备》 CAS 2013年第11期82-84,共3页 Chinese Medical Equipment Journal
关键词 血糖仪 质量控制 检测 blood glucose meter quality control test
  • 相关文献

参考文献10

二级参考文献12

  • 1孙虹,王凡,孙鷖,赵崇吉,张昱,牛华.透射和散射比浊法测定血清免疫球蛋白的评价[J].临床检验杂志,2005,23(3):189-191. 被引量:21
  • 2LANDT M.Glyceraldehyde preserves of glucose concentration in whole blood specimes[J].Clin Chem,2000,46(8Pt1):1144-1149.
  • 3THORNALLEY P J,STERN A.The effect of glyceraldehyde on red cells,haemoglobin status,oxidative metabolism and glyoolysis[J].Biochim Biophys Acta,1984,804(3):308-323.
  • 4ASTLES R,WILLIAMS C P,SEDOR F.Stability of plasma lactate in vitro in the presence of antiglycolytic agents[J].Clin Chem,1994,40(7Pt1):1327-1330.
  • 5The National Committee for Clinical Laboratory Standards. Method Comparison and Bias Estimation Using Patient Samples Approved Guideline [ S ]. NCCLS, EP9-A, 1995.
  • 6Porakishvili N, Fordham J L, Charrel M, et al. A low budget luminometer for sensitive chemiluminescent immunoassays [ J ]. Immunol Methods,2000,234 (1-2) :35-42.
  • 7Konig B,Ivankovic B,Schleff R,et al. Liaison free PSA-an automated chemiluminescent immunoassay for the determination of free prostate specific antigen[J]. Anticancer Res,1999,19(4A):2763-2765.
  • 8步怀恩,王泓午,李戈,尹立群,沈永明,赵铁牛.快速血糖仪与全自动生化分析仪测定血糖的Meta分析[J].天津医药,2008,36(2):119-121. 被引量:21
  • 9王海英,王凯锋.快速血糖仪与全自动生化分析仪血糖检测结果分析[J].中国误诊学杂志,2009(33):8148-8149. 被引量:14
  • 10朱春花,王静.两种抗凝剂抑制血糖分解效果分析[J].湘潭师范学院学报(自然科学版),2003,25(2):109-110. 被引量:5

共引文献56

同被引文献32

引证文献3

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部