摘要
通过采用不同类型的坩埚进行自动工业分析仪挥发分测试,检测坩埚对挥发分测试结果的影响。检测结果表明:坩埚质量相差在±1.000 0g范围,坩埚与盖应严密配合,且试样质量相差控制在±0.100 00g范围,以确保测试结果满足国标要求。
By using different types of crucible to do volatile test through automatic industrial analyzer,how the crucible affected the results could be tested.The testing results showed that:if the weight differences of crucibles were in the range of±1.000 0g,the cover of crucibl and crucible was matched well,and the weight differences were in the range of±0.100 00g,the testing results could meet the national standards.