摘要
利用钨丝特异性捕获镉(Cd)消除基体干扰,泡沫碳材质电热蒸发器串联原子荧光光谱仪,建立直接进样测定饮料痕量Cd的方法。仪器条件为:泡沫碳材质样品舟进样(10μL),采用灰化(矿泉水100W保持30s,其他饮料基质100W保持50s)与蒸发(70W保持30s)过程分离,Ar-H2(9:1)载气流速为600mL/min。结果表明:方法具有良好的线性,线性范围为0-5ng,相关系数r在0.998以上;10种饮料产品的加标回收率为92.2%~111.8%;方法检出限为0.04gg/L,相对标准偏差(RSD)在10%以内。本方法简单、灵敏、快速(包括样品前处理的分析时间不超过3min),适用于饮料产品痕量Cd的现场快速检测。
A novel direct sampling method to determine cadmium in beverages by solid sampling electrothermal vaporization atomic fluorescence spectrometry (SS-ETV-AFS) was established, where a tungsten coil as a Cd trap was used to remove matrix interferences and foam graphite as an electrothermal vaporizer and a sample boat. In the present study, 10 μL sample was loaded onto a sample boat, and then vaporized with 70 W for 30 s after being ashed with 100 W for 30 s (natural spring water) or 50 s (beverages), using Ar/H2(9:1) as the carrier gas with a flow rate of 600 mL/min. Results showed that the calibration curve for Cd had good linearity (from 0 to 5 ng), and a correlation coefficient (r) was more than 0.998; the recovery rates for Cd in 10 kinds of beverages were 92.2%-111.8%. The detection limit was 0.04 μg/L, and the relative standard deviations (RSD) were less than 10%. This method is simple, sensitive, fast (the whole analytical process including sample pretreatment can be completed within 3 min) and suitable for in-field determination of ultra-trace amounts of Cd in beverages.
出处
《食品科学》
EI
CAS
CSCD
北大核心
2013年第24期131-134,共4页
Food Science
基金
国家重大科学仪器设备开发专项(2011YQ14014904)
中央级公益性科研院所基本科研业务费专项(2012ZL014)
关键词
直接进样
钨丝捕获
电热蒸发
原子荧光光谱
饮料
痕量镉
direct sampling: tungsten coil trapping: electrothermal vaporization (ETV)
atomic fluorescencespectrometry (AFS)
beverages
ultra-trace cadmium