期刊文献+

用极差检验方案验证两个检测系统天冬氨酸氨基转移酶测定结果的可比性

Comparability of the Aspartate Aminotransferase Results Analyzed by Two Detection Systems and Verified through Range Test
下载PDF
导出
摘要 目的用极差检验可比性方案验证两台生化分析仪检测天冬氨酸氨基转移酶(AST)测定结果间的可比性。方法参照极差检验可比性方案,选用两份合适浓度的标本,对AST在两台生化分析仪上检测结果之间的可比性进行研究,判断不同检测系统之间是否具有可比性。结果基于生物学变异可比性的评价标准,两个检测系统测定的AST结果所得极差值小于设定的临界差值。结论两个检测系统AST测定结果具有可比性;极差检验可比性方案是一种简单、实用性很强的可比性验证新方案。 Objective To explore the comparability of the AST results analyzed on two biochemistry analyzers and verified through range test. Methods According to the range test comparability scheme, two samples of different AST concentration were selected and analyzed respectively on two biochemistry analyzers to assess the compara- bility of the detection systems. Results Based on the evaluation standards of the comparable biological variability, the range value obtained by the two detection systems was under acceptable criteria. Conclusion The AST results analyzed by the two detection systems were absolutely comparable, and the range test compara- bility scheme is simple and eminently practical.
出处 《标记免疫分析与临床》 CAS 2013年第6期440-442,共3页 Labeled Immunoassays and Clinical Medicine
关键词 极差检验 生化分析仪 可比性 AST Range test Biochemistry analyzer Comparability AST
  • 相关文献

参考文献4

二级参考文献16

  • 1王霞,江虹,贾劲,周静,廖娟.应用病人标本对STA-COMPACT与STA-R全自动血凝分析仪进行比对[J].华西医学,2006,21(4):767-768. 被引量:7
  • 2张秀明,郑松柏,孙蕾,柯培锋,马骥,张文,林莲英.应用Westgard方法评价决定图判断生化检测系统性能的可接受性[J].中华检验医学杂志,2007,30(1):86-90. 被引量:73
  • 3魏吴 丛玉隆.医学实验室质量管理与认可指南[M].北京:中国计量出版社,2004.27-80.
  • 4朱文元,廖昆灵,刘芹.不同凝血分析仪检测纤维蛋白原结果的一致性评价[J].检验医学与临床,2012,9(4):479.
  • 5NCCLS. EP9-A2 Method comparison and bias estimation using patients samples.- approved guideline E S. 2nd ed. Wayne, PA, USA.. NCCLS, 2006.
  • 6卫生部临床检验中心,中国医院协会临床检验管理专业委员会.《医疗机构临床实验室管理办法》及配套文件[S].北京:卫生部临床检验中心,2008.
  • 7National Committee for Clinical Laboratory Standards. Method comparisonand bias estimation using patient samples [S]. Ap- proved Guideline ,2nd ed ,EP9-A2,2002.
  • 8府伟灵,徐克前.临床生物化学检验[M].第5版.北京:人民卫生出版社,2012.1:15.
  • 9Henry C, Royden NR. Early History of Eastman Kodak ekat- achem slides[J]. Clin Chem, 1997,63 (9) : 647.
  • 10颜存粮,彭黎明,黄海雄.不同血凝分析仪检测结果的一致性研究[J].中华检验医学杂志,2008,31(1):100-103. 被引量:15