期刊文献+

正常实验猪心肌灌注成像中伪影探讨 被引量:1

The Endocardial “Dark Rim” Artifact in MRI Myocardial Perfusion Imaging of Normal Pigs
下载PDF
导出
摘要 目的:分析正常实验猪MR心肌灌注成像中伪影形成的影响因素,并总结伪影表现。方法:选择11头正常家猪进行静息及潘生丁负荷MR心肌灌注成像检查。所有实验猪均经冠状动脉造影证实无冠状动脉病变。依据心肌节段伪影有无与伪影出现时相(磁敏感度)、心脏收缩运动、相位编码方向及层厚(through-plane空间分辨率)之间的关系进行两分类Binary Logistic回归分析。结果:磁敏感度、心脏收缩运动及相位编码方向三者与心肌灌注成像伪影的形成存在一定关系,其中以磁敏感度影响最大(OR值=5.735)。伪影多出现在心腔信号最强而心肌开始强化的时刻,即受磁敏感度影响,呈一过性表现。层厚与伪影形成关系不大。结论:MR心肌灌注中出现的心内膜下伪影受多种因素影响,以磁敏感度影响最明显。认识该伪影表现对鉴别诊断及半定量分析大有裨益。 Purpose: To analyze causes and appearance of the endocardial "dark rim" artifact in MR firstpass myocardial perfusion imaging. Methods: First-pass contrast-enhanced MR imaging was performed at rest and after the infusion of dipyridamole in 11 healthy pigs that were proved to be normal in the coronary angiography before MR scans. The association between the band artifact and the influence factors including the susceptibility effect from the gadolinium bolus, cardiac motion, the direction of the phase-encoding (inplane spatial resolution) and slice thickness (through-plane spatial resolution) was analyzed. Binary logistic regression was used to evaluate the relationship of them. Results: The dark rim artifact was due to the combination of susceptibility effect, cardiac motion and in-plane spatial resolution. Among them, the influence of susceptibility effect was the largest, whose OR (odds ratio) was 5.735. The "dark rim" artifact often raised in tandem with the arrival of the bolus of gadolinium in the left ventricle. It typically preceded myocardial tissue uptake and was transient in nature. However, the effect of slice thickness was limited. Conclusion: The endocardial "dark band or rim" artifact is due to many factors, while the susceptibility effect plays the dominating role. The recognition of the appearance of the artifact will be helpful for differentiating it from the true perfusion defects.
出处 《中国医学计算机成像杂志》 CSCD 北大核心 2013年第6期563-566,共4页 Chinese Computed Medical Imaging
关键词 伪影 心肌灌注 磁共振成像 Artifacts Myocardial perfusion imaging Magnetic resonance imaging
  • 相关文献

参考文献13

  • 1Barkhausen J,Hunold P,Jochims M. Imaging of myocardial perfusion with magnetic resonance[J].{H}Journal of Magnetic Resonance Imaging,2004.750-757.
  • 2Storey P,Chen Q,Li W. Band artifacts due to bulk motion[J].{H}Magnetic Resonance in Medicine,2002.1028-1036.
  • 3Bella ED,Parker DL,Sinusas AJ. On the dark rim artifact in dynamic contrast-enhanced MRI myocardial perfusion studies[J].{H}Magnetic Resonance in Medicine,2005.1295-1299.
  • 4Arai AE. Magnetic resonance first-pass myocardial perfusion imaging[J].{H}Topics in Magnetic Resonance Imaging,2000.383-398.
  • 5Elkington AG,Gatehouse PD,Cannell TM. Comparison of hybrid echo-planar imaging and FLASH myocardial perfusion cardiovascular MR imaging[J].{H}RADIOLOGY,2005.237-243.
  • 6Saeed M,Wendland M,Yu K. Dual effects of gadodiamide injection in depiction of the region of myocardial ischemia[J].{H}Journal of Magnetic Resonance Imaging,1993.21-29.
  • 7Schwitter J. Myocardial Perfusion[J].{H}Journal of Magnetic Resonance Imaging,2006.953-963.
  • 8Su MY,Yang KC,Wu CC. First-pass myocardial perfusion cardiovascular magnetic resonance at 3 Tesla[J].{H}JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE,2007.633-644.
  • 9Sodickson DK,Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH):fast imaging with radiofrequency coil arrays[J].{H}Magnetic Resonance in Medicine,1997.591-603.
  • 10Pruessmann KP,Weiger M,Scheidegger MB. SENSE:sensitivity encoding for fast MRI[J].{H}Magnetic Resonance in Medicine,1999.952-962.

二级参考文献7

  • 1Gelb A. Parameter optimization and reduction of round off error for the Gegenbauer reconstruction method [ J ]. Journal of Scientific Computing, June 2004, 20(3 ) : 433-459.
  • 2Gelb A, Tadmor E. Detection of edges in spectral data[ J]. Applied and Computational Harmonic Analysis, 1999, 7( 1 ) : 101-135.
  • 3Gelb A, Tadmor E. Detection of edges in spectral data Ⅱ. nonlinear enhancement[ J]. Society for Industrial and Applied Mathematics,2000, 38(4) : 1389 N 1408.
  • 4Archibald R, Gelb A. A Method to reduce the Gibbs ringing artifact in MRI scans while keeping tissue boundary integrity [ J ]. IEEE Transactions on Medical Imaging, 2002, 21(4) : 100 -113.
  • 5Gottlieb D, Shu C W. On the Gibbs phenomenon and its resolution[J]. SIAM Review, 1997, 39(4): 644-668.
  • 6Shizgal B D, Jung J H. Towards the resolution of the Gibbs phenomena[J]. Journal of Computational and Applied Mathematics,2003, 161(1) : 41-65.
  • 7Jung J H, Shizgal B D. Generalization of the inverse polynomial reconstruction method in the resolution of the Gibbs phenomenon [ J ].Journal of Computational and Applied Mathematics, 2004, 172 ( 1 ) :131-151.

共引文献1

同被引文献6

引证文献1

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...