MOSFET探测器在全身放疗质量控制中的应用 被引量:1

Application of MOSFET Detector in the Quality Control of Total Body Irradiation
下载PDF
导出
摘要 目的探讨MOSFET探测器在全身放疗质量控制中的价值。方法在源轴距360 cm处使用电离室对MOSFET探测器进行标定,同时根据体模入射面、出射面与中间层面的剂量换算关系,利用标定后的MOSFET探测器检测行全身放疗患者的辐射剂量,控制并减少相应误差。结果 10个MOSFET探测器中,有7个最大偏差均<3%,另3个经再次重新标定后,最大偏差亦<3%,均可用于临床测量。结论在全身放疗中应用MOSFET探测器能够起到监测治疗剂量、控制误差的作用。 Objective To investigate the application value of MOSFET detector in the quality control of total body irradiation. Methods To demarcate the radiation dose of MOSFET detector with ionization chamber at 360 cm SAD. To detect the radiation dose of patients during total body irradiation by using demarcated MOSFET detector in order to control and reduce measuring errors according to the conversion relationship among incident surface, exit surface and intermediate layer of the model. Results Among all MOSFET detectors (10 cases), the maximum dose deviation (〈3%) can meet clinical requirements in 7 cases and the maximum dose deviation (〈3%) also can meet clinical requirements in 3 cases which are demarcated again. Conclusion The radiation dose can be monitored and measuring errors can be controlled with the application of MOSFET detector in total body irradiation.
出处 《中国医疗设备》 2014年第1期117-118,140,共3页 China Medical Devices
关键词 MOSFET探测器 全身放疗 质量控制 X射线 MOSFET detector total body irradiation quality control X-ray
  • 相关文献

参考文献7

二级参考文献25

  • 1Vassiliki Vlachopoulou,Georgia Malatara,Harry Delis,Kiki Theodorou,Dimitrios Kardamakis,George Panayiotakis.Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET[J].World Journal of Radiology,2010,2(11):434-439. 被引量:1
  • 2吕庆文,陈超敏,周凌宏,李树祥,林世寅.精确放射治疗技术的现状与发展[J].中国医学物理学杂志,2005,22(1):373-375. 被引量:10
  • 3谷铣之,殷蔚伯,刘泰福,等.肿瘤放射治疗学.第2版.北京:北京医科大学、中国协和医科大学联和出版社,1998.209.
  • 4Soubra M,Cygler J,MacKay,GF.Evaluation of a dual bias dual metaloxide-silicon semiconductor field effect transistor detector as a radiation dosimeter.Med Phys,1994,21:567-572.
  • 5Scalchi P,Francescon P.Calibration of a mosfet detection system for 6MV in vivo dosimetry.Int Radiat Oncol Biol Phys,1998,40:987-993.
  • 6Chuang C,Verhey L,Xia P.Investigation of the use of mosfet for clinical IMRT dosimetric verification.Med Phys,2002,29:1109-1115.
  • 7Sessions JB,Roshau JN,Tressler MA,et al.Comparisons of point and average organ dose within an anthropomorphic physical phantom and a computational model of the newborn patient.Med Phys,2002,29:1080-1089.
  • 8Ribas M, Jomet N, Eudaldo T, et al. Midplane dose deterrninafion during total body irradiation using in vivo dosimetry.Racliother Oncol, 1998, 49:91 -98.
  • 9Van Dam J. Correlation between temperature and dose rate dependence of semiconductor response influence of accumulated dose.Radiother Oncol, 1990, 19: 345.
  • 10张红志.放射治疗质量保证:新技术的的挑战[A].中华医学会放射肿瘤治疗学分会六届二次暨中国抗癌协会肿瘤放疗专业委员会二届二次学术会议论文集[c].2007.

共引文献22

同被引文献9

  • 1王军良,周振山,盛洪国,齐伟华.IAEA TRS398与TRS277应用于加速器输出量校准的比较[J].中国医学物理学杂志,2011,28(3):2594-2596. 被引量:11
  • 2American Association of Physicists in Medicine.TG 21.A protocol for the determination of absorbed dose from high-energy photon and electron beams[J].Med Phys, 1983,10(6):741- 771.
  • 3American Association of Physicists in Medicine.TG 51 .Protocol for clinical reference dosimetry of high-energy photon and electron beam[J].Med Phys, 1999,26(9): 1847-1870.
  • 4World Health Organisation.Quality assurance inradiotherapy[S]. 1988.
  • 5International Atomic Energy Agency.TRS 277.Absorbed dose determination in photon and electron beams:aninternational code of practice[S]. 1997.
  • 6International Atomic Energy Agency.TRS 398.Absorbed dose determination in external beam radiotherapy:An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water[S].2000.
  • 7Andreo P, Saiful M,Huq MS,et al.Protocols for the dosimetry of high-energy photon and electron beams:a comparison of the IAEA TRS-398 and previous international Codes of Practice[J].Phys Med Biol,2002,47(17):3033-3053.
  • 8胡杰,陶建民,张莹,余根培,张颖.放射治疗中绝对剂量的测量[J].中国医疗器械杂志,2011,35(3):225-228. 被引量:10
  • 9文婷,黎静,张晋建,余咏兰,尹智宇,靳怀志,王雅冰,叶柳青.螺旋断层放疗的日常质量控制与质量保证[J].中国医疗设备,2013,28(7):50-52. 被引量:8

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部