XE-5000血细胞分析仪计数有核红细胞的性能评价 被引量:3

Research on Enumerating Nucleated Red Cells by Sysmex XE-5000
下载PDF
导出
摘要 目的探讨SysmexXE.5000血细胞分析仪自动计数外周血有核红细胞(NRBC)的方法学特点,评价其临床应用价值。方法评析其重复性、线性范围、携带污染率,并对697份静脉血标本作常规测定,评价其对NRBC的报警、定量测定,并与手工涂片法进行对照。结果XE.5000血液分析仪测定NRBC的重复性较好,线性范围较广,携带污染率较小。以显微镜复检为金标准,XE.5000与显微镜计数法计数外周血中NRBC数量有极好的相关性(r=0.997),其有核红细胞报警的灵敏度为100%,特异度为85.1%。结论SysmexXE.5000血细胞分析仪可灵敏、准确、精密地自动计数外周血中NRBC,适合临床检测NRBC。 Objective To explore the methodology characteristics of automated method of enumerating nucleated red blood cells(NRBC) in peripheral blood on Sysmex XE-5000,and to evaluate its clinical application value. Methods Precision,linearity and carryover were performed by assaying the differential NRBC samples. The results of 697 blood samples measured by Sysmex XE-5000 were compared with manual microscope method. Result The System XE-5000 results showed excellent precision, linearity and stability. As a gold standard of manual method, the correlation was excellent between automated method and manual method(r=0.997). The specificity for NRBC was 85.1%, and sensitivity was 100%. Conclusion NRBC counting by XE-5000 was sensitive, accurate and reliable, which is suitable for clinical detection of NRBC.
出处 《中国血液流变学杂志》 CAS 2013年第3期544-546,共3页 Chinese Journal of Hemorheology
关键词 有核红细胞 血细胞分析仪 nucleated red blood cell blood cell analyzer
  • 相关文献

参考文献10

  • 1张之南;杨天楹;郝玉书.血液病学[M]{H}北京:人民卫生出版社,20031670-1675.
  • 2张之南;沈悌.血液病诊断及疗效标准[M]{H}北京:科学出版社,1999103-175.
  • 3National Committee for Clinical Laboratory Standards. Reference leukocyte differential count(proportional)and evaluation of instrument methods:approved standard[S].USA,1992.
  • 4de Keijzer MH,van der Meer W. Automated counting of nucleated red blood cells in blood samples of newborns[J].{H}Clinical Laboratory Hematology,2002,(06):343-345.
  • 5Walters J,Garrity P. Performance evaluation of the Sysmex XE-2100 hematology analyzer[J].Lab Hemat,2000,(02):83-92.
  • 6Gulati G,Behling E,Kocher W. An evaluation of the performance of Sysmex XE-2100 in enumerating nucleated red cells in peripheral blood[J].{H}ARCHIVES OF PATHOLOGY AND LABORATORY MEDICINE,2007,(07):1077-1083.
  • 7Wang FS,Itose Y,Tsuji T. Development and clinical application of nucleated red blood cell counting and staging on the automated haematology analyser XE-2100[J].{H}Clinical Laboratory Hematology,2003,(01):17-23.
  • 8李建英.血细胞分析仪自动计数外周血有核红细胞的应用研究[J].中华检验医学杂志,2005,28(6):606-609. 被引量:23
  • 9Kratz A,Maloum K,O'Malley C. Enumeration of nucleated red blood cells with the ADVIA 2120 Hematology System:an International Multicenter Clinical Trial[J].Lab Hematol,2006,(02):63-70.
  • 10王欣,滕娟,郝少丽.血细胞分析仪计数幼稚红细胞的准确性评价[J].中华检验医学杂志,2007,30(9):1056-1057. 被引量:7

二级参考文献10

  • 1杨一芬,曹虹,董存岩.Sysmex XE-2100检测网织红细胞计数评价[J].中国现代医学杂志,2005,15(9):1404-1406. 被引量:4
  • 2熊立凡.临床检验基础[M].北京:人民卫生出版社,2004.25.
  • 3De Keijer MH, van Der Meer W. Automated counting of nucleated red blood cells in blood samples of newborns. Clin Lab Haem ,2002,24:343-345.
  • 4Wang FS, Itose Y, Tsuji T, et al. Development and clinical application of nucleated red blood cell counting and staging on the automated haematology analyzer XE-2100. Clin Lab Haem, 2003,25:17-23.
  • 5Walter J, Garrity P. Performance evaluation of the Sysmex XE-2100 hematology analyzer. Lab Hemat ,2000, 6:83-92.
  • 6De Keijer MH, van Der Meer W. Automated counting of nucleated red blood cells in blood samples of newborns. Clin Lab Haem, 2002,24 :343-344.
  • 7Ruzicka K, Veitl M, Thalhammer-Scherrer R, et al. The new hematology analyzer Sysmex XE-2100: performance evaluation of a novel white blood cell differential technology. Arch Pathol Lab Med, 2001,125:391-396.
  • 8Walter J, Carrity PN. Performance evaluation of the sysmex XE- 2100 hematology analyzer. Lab Hermit, 2000, 6:83-92.
  • 9中华医学会检验分会.临床血液和尿液分析专题研讨会纪要[J].中华医学检验杂志,1995,19:122-125.
  • 10顾可梁.使用血液细胞分析仪有关问题解答[J].临床检验杂志,2000,18(1):58-59. 被引量:36

共引文献26

同被引文献21

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部