摘要
采用连续流动分析仪测定老挝钾盐矿中的钾含量,并与四苯硼钠重量法进行了比较。结果表明,该方法相关系数达到1.0000,RSD在0.61%~0.88%之间,加标回收率在97.6%~102.4%之间。
The potassium in Laos Potash Mine was determined by Continuous Flow Analyzer,and compared with Sodium Tetraphenylboron. The results showed that the analyzer was able to determine accurately and efficiently the potassium content in Laos Potash Mine. Its coefficient of correlation reached 1. 0000,RSD was at 0. 61% ~ 0. 88% and the recovery of spiked standard at 97. 6% ~ 102. 4%.
出处
《云南化工》
CAS
2013年第6期87-89,共3页
Yunnan Chemical Technology
关键词
老挝
钾盐矿
钾
连续流动分析仪
四苯硼钠
Laos potash
potash mine
potassium
continuous flow analyzer
sodium tetraphenylboron