摘要
目的研究环氧乙烷的灭菌效果及残留量。方法用微生物学法和气相色谱法测定。结果及结论用环氧乙烷灭菌的塑料瓶内塞能达到无菌要求,环氧乙烷残留量符合要求。在严格按要求操作的前提下,生产上可用环氧乙烷灭菌法为不耐热内包材灭菌。
Objective To study the sterilization effect and residual level of ethylene oxide. Methods The microbiological sterility test and gas chromatography are adopted. Results and Conclusion Inner stoppers of plastic bottles after sterilization with ethylene oxide can meet the aseptic requirement and the residual level of ethylene oxide can also meet the requirement. Ethylene oxide can be used in the sterilization of thermolabile inner packaging material in the production process.
出处
《食品与药品》
CAS
2014年第1期41-43,共3页
Food and Drug
关键词
环氧乙烷
灭菌
残留量
ethylene oxide
sterilization
residual level