摘要 目的探讨全自动血细胞分析仪异常白细胞散点图在仪器故障及疾病诊断过程中的价值。方法对工作中出现的6例异常白细胞散点图进行分析。结果附图1、2反映了白细胞散点图与仪器试剂故障的关系;附图3反映了白细胞散点图与白血病诊断的关系;附图4反映了白细胞与疟疾诊断的关系;附图5反映了白细胞与EVENS综合症诊断的关系;附图6反映患者粒生素治疗后白细胞急性升高的图形。结论全自动血细胞分析仪散点图能为仪器故障的早期发现和疾病的早期诊断和疗效观察提供有价值的信息。 Objective To investigate the value of abnormal white blood cell scatter of automatic blood cell analyzer in the equip- ment failure and disease diagnosis. Methods Six cases of abnormal white blood cell scatter plot were analyzed. Results The first and second graphics reflected the relationship between leukocyte scatter plots and the instrument or reagent failure;the third graph- ic reflected the Felatlonship between leukocyte scatter plot and the diagnosis of leukemia; the fouth graphic reflected the relationship between leukocyte scatter plot and malaria diagnosis;the fifth graphic reflected the relationship between leukocyte scatter plot and EVENS syndrome diagnosis,the sixth graphic reflected the white blood cells rise after granulocyte colony stimulating factor treat- ment. Conclusion The automatic blood cell analyzer can provide powerful information for early detection of the equipment failure, early diagnosis of diseases and curative effect.
作者 王朝辉 仲华
出处 《国际检验医学杂志》 CAS 2014年第3期338-339,共2页 International Journal of Laboratory Medicine
关键词 血细胞分析仪 白细胞散点图 白细胞计数 blood cell analyzer white blood cell scatter plots leukocyte count