期刊文献+

时间分辨荧光免疫法风疹病毒IgG抗体定量测定试剂盒的研制与性能评价 被引量:8

Development and performance evaluation of time-resolved fluorescence immunoassay quantitative kit on determination of rubella virus IgG antibody
下载PDF
导出
摘要 目的建立风疹病毒(RV)IgG抗体时间分辨荧光免疫法(TrFIA)并研制其试剂盒。方法采用RV抗原作为包被抗原,铕标记羊抗人IgG作为示踪物,配制以β一萘甲酰三氟丙酮为主要成分的增强液,应用TrFIA法测定RVIgG,并对自研试剂盒的最低检测限、准确度、线性、重复性、阴性参考品符合率、阳性参考品符合率、热稳定性进行评价,与同类试剂盒进行方法学比对试验,比对试验定量结果的等效性采用线性回归分析。结果自研试剂盒的最低检测限、重复性、阴性参考品符合率、阳性参考品符合率均能达到国家检定标准;检测国际参考品(10IU/mL)或企业校准品,其实测值与理论值相对偏差均在正负20.0%以内;在2.0~256.0IU/mL范围内,线性相关系数不低于0.9900;于37℃恒温箱放置6d后,检测性能无明显改变;临床研究评价一致程度百分比可达100%。结论自研试剂盒灵敏度高、特异性强、精密度好、准确度高、线性范围宽,与同类产品检测结果相关性好,能满足临床需要。 Objective To establish the rubella virus (RV) IgG antibody time-resolved fluorescence immunoassay(TrFIA) and the development of test kit. Methods RV antigen was taken as coating antigen,europium-labeled goat anti-human IgG was taken as a tracer, we constructed a TrFIA kit and the self-developed kit was evaluated by testing the lowest limit of detection,accuracy, line- arity,and repeatability. Results The kit's detection limit,repeatability are in line with the rate of negative reference materials. The measured value is highly proportioned to the theoretical value;within 2.0-256.0 IU / mL range, the linear correlation coefficient of higher than 0. 990 0. Conclusion The self-veloped detection kits provide satisfactory sensitivity,specificity,precision,and accura- cy,and it can be of help in clinical detection.
出处 《国际检验医学杂志》 CAS 2014年第4期472-474,共3页 International Journal of Laboratory Medicine
基金 广东省科技计划项目(2010B050300002) 国家高技术研究发展("863"计划)项目(2011AA02A112)
关键词 时间分辨荧光免疫法 风疹病毒 IGG抗体 time-resolved fluorescence immunoassay rubella virus IgG antibodies
  • 相关文献

参考文献9

二级参考文献74

共引文献32

同被引文献64

引证文献8

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部