旋转照射剂量测量仪(ArcCHECK)在螺旋断层放疗计划剂量验证中的应用 被引量:12

Application of ArcCHECK in Dose Verification of Tomotherapy Planning
原文传递
导出
摘要 目的:研究旋转照射剂量测量仪(ArcCHECK)在螺旋断层放疗(HT)计划剂量验证中的应用,寻求一种用于临床上的剂量验证方法。方法:采用ArcCHECK对36例患者螺旋断层放疗(HT)计划进行剂量验证,用SNC Patients软件进行剂量分布γ值分析,评估计划的剂量验证结果及实施的可行性。结果:36例患者HT计划剂量验证结果表明,采用Gamma法(3mm/3%)分析,γ≤1像素点平均通过率为99.7%±0.6%,γ通过率都大于95%。结论:ArcCHECK作为螺旋断层放疗计划的剂量验证稳定,可用于HT计划剂量分布验证。 Objective: To evaluate the dosimetric characteristics of ArcCHECK application in dose verification of Helical Tomotherapy planning. Methods: Dose verifications of Tomotherapy planning of 36 patients were performed using ArcCHECK. The γ analysis of dose distribution was conducted with SNC Patients software. Results : The dose distribution measured by AreCHECK was well consistent with that calculated by Tomotherapy DQA station. In comparison of Arc- CHECK measured dose versus DQA calculated with 3mm/3% ~ criteria using SNC Patient software, the average pass rate of pixels of 36 cases with γ≤ 1 was 99.7% +0.6%. The -γ pass rates of the 36 cases were over 95%. Conclusion: Arc- CHECK has been successfully tested for dose verification of HT planning, and offers an accurate and convenient rotational dosimetric solution.
出处 《肿瘤预防与治疗》 2014年第1期20-23,共4页 Journal of Cancer Control And Treatment
关键词 螺旋断层放疗 剂量验证 ArcCHECK γ分析 Tomotherapy Dose Verification ArcCHECK Gamma Analysis
  • 相关文献

参考文献8

  • 1Mackie TR,Holmes T,Swerdloff S. Tomotherapy:A new cowept for the delivery of dynamic conformal radiotherapy[J].Medical Physics,1993,(06):1709-1719.
  • 2Mackie TR,Balog J,Ruchala K. Tomotherapy[J].Seminars in Radiation Oncology,1999,(01):108-117.
  • 3徐寿平,邓小武,戴相昆,王连元,解传滨,葛瑞刚,沙翔燕.螺旋断层放疗系统调强放疗验证[J].中华放射肿瘤学杂志,2008,17(5):395-397. 被引量:17
  • 4Xu SP,Deng XW,Dai XK. Quality assurance of helical tomotherapy intensity modulated radiation therapy[J].APCMBE IFMBE Proceedings,2008,(05):447-450.
  • 5徐寿平,解传滨,鞠忠建,戴相昆,郭妍妍,巩汉顺,王连元.用二维电离室阵列对螺旋断层治疗的调强计划进行剂量验证[J].中华放射肿瘤学杂志,2009,18(3):233-236. 被引量:9
  • 6Low DA,Harms WB,Mutic S. A technique for the quantitative evaluation of dose distributions[J].Medical Physics,1998,(05):656-661.
  • 7Low DA,Dempsey JF. Evaluation of the gamma dose distribution comparison method[J].Medical Physics,2003,(09):2455-2464.
  • 8Van Dyke J,Bamett RB,Cygler JE. Commissioning and quality assurance of treatment planning computers[J].International Journal of Radiation Oncology ? Biology ? Physics,1993,(02):261-273.

二级参考文献16

  • 1戴建荣,胡逸民,张红志,关莹,张可,王闯.针对患者调强放射治疗计划的剂量学验证[J].中华放射肿瘤学杂志,2004,13(3):229-233. 被引量:65
  • 2Mackie TR, Holmes T, Swerdolff S.et al. Tomotherapy : a new concept for the delivery of conformal radiotherapy. Med Phys, 1993, 20 : 1709-1719.
  • 3Mackie TR,Balog J,Ruchala K,et al. Tomotherapy. Semin Radiat Oncol, 1999,9 : 108-117.
  • 4Yang JN, Mackie TR, Reckwerdt P, et al. An investigation of tomotherapy beam delivery. Med Phys.1997,24 425-436.
  • 5Balog J. Tomotherapy dosimetry and the tomotherapy workbench. Madison: University of Wisconsin,1998.
  • 6Ezzell GA,Galvin JM,Low D,et al. Guidance document on delively,treatment planning, and clinical implementation of IMRT: report of the IMRT subconmlittee of the AAPM radiation therapy committee. Med Phys,2003,30:2089-2115.
  • 7Dong L, Antolak J,Sa ehpour M,et al. Patient-specific point dose measurement for IMRT monitor unit verification. Int J Radiat Oncol Biol Phys,2003,56:867-877.
  • 8Olivera GH, Shepard DM, Reckwerdl PJ. et al. Maximum likelihood as a common computational framework in tomotherapy. Phys Med Biol,1998 ,43 :3277-3294.
  • 9Low DA, Dempsey JF. Evaluation of the gamma close distribution comparison metbod. Med Phys,2003,30 : 2455-2464.
  • 10Woo M, Nico A. Using a matrix detector for rotational delivery QA. Med Phys ,2007 ,34 :2502.

共引文献22

同被引文献90

  • 1马林,王连元.周桂霞.肿瘤断层放射治疗[M].成都:四川科学技术出版社,2010:55.
  • 2中华人民共和国卫生部.GBZ126-2011电子加速器放射治疗放射防护要求[S].北京:人民卫生出版社,2011.
  • 3Nguyen NP,Vock J,Sroka T,et al.Feasibility of image-guided radiotherapy based on tomotherapy for the treatment of locally advanced anal carcinoma[J].Anticancer Res,2011,31(12): 4393-4396.
  • 4Balog J,Soisson E.Helical tomotherapy quality assurance[J].Int J Radiat Oncol Biol Phys,2008,71 Suppl 1: S113-117.
  • 5Giraud P,Kantor G,Yassa M,et al.Two-year clinical experience with tomotherapy: the French National cancer institute project on implementing new technology[J].Cancer Invest,2011,29(8): 557-563.
  • 6International Commission on Radiological Protection.ICRP publication 112.A report of preventing accidental exposures from new external beam radiation therapy technologies[J].New York: Elsevier,2009.
  • 7Langen KM,Papanikolaou N,Balog J,et al.QA for helical tomotherapy: report of the AAPM Task Group 148[J].Med Phys,2010,37(9): 4817-4853.
  • 8Broggi S,Cattaneo GM,Molinelli S,et al.Results of a two-year quality control program for a helical tomotherapy unit[J].Radiother Oncol,2008,86(2): 231-241.
  • 9Staton RJ,Langen KM,Kupelian PA,et al.Dosimetric effects of rotational output variation and X-ray target degradation on helical tomotherapy plans[J].Med Phys,2009,36(7): 2881-2888.
  • 10Darby S, McGale P, Correa C. Early Breast Cancer Trialists' Collabora- tive Group (EBCTCG). Effect of radiotherapy after breast-conserving surgery on 10-year recurrence and 15-year breast cancer death: meta-analysis of individual patient data for 10, 801 women in 17 rand-omised trials[J]. Lancet, 2011,378(9804): 1707-1716.

引证文献12

二级引证文献29

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部