目的:通过对两台不同全自动血沉分析仪检测结果的比对,实现同一实验室不同仪器间检测结果的一致性。方法:随机抽取50例标本,分别用两台仪器检测血沉,对数据进行统计学分析。结果:两台仪器血沉检测结果的相关性欠佳(r=0.92),结果差异具有统计学意义(P<0.05)。结论:两台血沉仪检测结果不一致,应对仪器参数重新设定。
Objective: To compare the consistency of two different automated analyzers in detecting erythrocyte sedimentation rate. Methods: Fifty patients were selected randomly and detected erythrocyte sedimentation rate with two different automated erythrocyte sedimentation rate analyzer and in which the data were statistically analyzed and compared. Results: The consistency of two instruments was poor (r=0.92) in which the result of the existence was significant differences (P〈0.05). Conclusion: Two blood sedimentation instrument test results are not consistency in which the instrument parameters should be reseted.