基于图像质量增强的大螺距重建技术及其应用

Application of reconstruction algorithm based on image quality enhancement
下载PDF
导出
摘要 目的评价基于大螺距原始投影数据的图像质量增强重建(IQE)技术的图像质量。方法①采集头颅标本原始数据,分别以传统方式和IQE方式进行重建并评价图像伪影。②选择60例于16排CT接受头颅检查患者,其中30例采用螺距1∶1.75,另30例采用螺距1∶1.375,管电压及管电流均为120kVp、200 mAs,重建层厚和重建间隔均为0.625mm。对1∶1.75螺距采样原始数据分别采用常规技术和IQE技术进行重建,评价不同螺距、不同重建方式图像质量的差异,并获取图像噪声标准差(SD)值进行定量分析。结果①头颅模型数据显示,采用IQE重建较传统重建图像中的风车状伪影明显减少。②临床图像数据显示,螺距为1∶1.75时,采用IQE重建图像的伪影明显少于传统技术重建图像;IQE图像质量优于传统重建(P<0.05)。传统重建技术和IQE重建图像的SD值分别为27.42±3.70和20.11±4.20(t=4.913,P<0.05)。结论采用IQE技术进行大螺距扫描可获得满意的图像质量,并降低图像噪声。 Objective To evaluate the image quality using image quality enhancement (IQE) algorithm based on projection data of high pitch scan. Methods (1)Original data of a aduIt head specimen was obtained. And a subjective evaluation of image artifacts was done. (2)Sixty patients who underwent head CT scanning were selected, and the parameters were set as follow: 30 patients were scanned with pitch 1 : 1.75, and the other with pitch 1 : 1. 375, 120 kVp, 200 mAs. Traditional technique and IQE were both used to reconstruct images. The image quality and standard deviation (SD) value were measured and compared, and the data were quantitatively analyzed. Results OSpecimen data: Windmill artifacts in images reconstructed with IQE were less than those reconstructed with traditional technique. (2)Clinical data.. With pitch 1 : 1.75, image artifact was lower with IQE reconstruction, and image quality using IQE was higher than that using traditional technique. SD value was 27.42±3.70 and 20.11±4.20 respectively for IQE and traditional technique (t=4. 913, P〈0.05). Conclusion IQE can be used to reconstruct images with high quality and less image noise.
作者 程传乐
出处 《中国介入影像与治疗学》 CSCD 2014年第3期188-191,共4页 Chinese Journal of Interventional Imaging and Therapy
关键词 体层摄影术 螺旋计算机 伪影 图像质量 Tomography, spiral computed Artifacts Image quality
  • 相关文献

参考文献9

  • 1Nakano S, Sakamoto H, Successful use of multi detector row computed tomography for detecting contra[atera[ hreast cancer. J Comput Assist Tomogr, 2011,35(1):148- 52.
  • 2Vikgren J, Johnsson AA, High-resolution computed tomography with 16-row MDCT: A comparison regarding visibility and motion artifacts of dose modulated thin slices and " step and shoot" ima- ges, Acta Radiol, 2008,49(7):755- 760.
  • 3刘鑫,郭启勇,林爱军,廖伟,苏惠群.低剂量MD-CT气管支气管树重建的动物实验研究[J].中国医学影像技术,2005,21(1):9-13. 被引量:10
  • 4刘学静,武乐斌,柳澄,王道萍,王丽君.多层螺旋CT门静脉造影术准直与螺距匹配及重建方法优化选择[J].中国医学影像技术,2004,20(12):1933-1936. 被引量:3
  • 5Hsieh J. Nonlinear Partial volume artifact correction in helical CT. IEEE Transactions on Nuclear Science, 1999, 46 (3):2035- 2044.
  • 6Kyriakou Y, Kachelriess M, Nnaup M, et al. Impact of the z-fly- ing focal spot on resolution and artifact behavior for a 64-slice spi ral CF scaane. Eur Radiol, 2006,16(6) :1206-1215.
  • 7Chrismer JA, Stierstorfer K, Primak AN. Evaluation of z-axis resolution and image noise for nonconstant velocity spiral CT data reconstructed using a weighted 3D filtered backprojection (WFBP) reconstruction algorithm. Med Phys, 2010, 37 (2) :897 906.
  • 8彭芸,李剑颖,马大庆.CT检查中低X射线剂量技术的应用和进展[J].中华放射学杂志,2008,42(10):1117-1120. 被引量:69
  • 9Kambadakone AR, Prakash P, Hahn PF, et al. Low-dose CT examinations in Crohffs disease: Impact on image quality, diag nostic performance, and radiation dose. AJR Am J Roentgenol, 2010, 195(1) :78-88.

二级参考文献68

  • 1柳澄,秦维昌.多层螺旋CT(一)[J].医学影像学杂志,2000,10(3):194-195. 被引量:42
  • 2金武男,杨香.螺旋CT三维重建对各肝段门静脉解剖结构的观察[J].中国医学影像技术,2003,19(6):692-695. 被引量:14
  • 3Colang JE, Killion JB, Vano E. Patient dose from CT: a literature review. Radiol Technol,2007,79 : 17-26.
  • 4Brenner D J, Hall EJ. Computed tomography:an increasing source of radiation exposure. N Engl J Med,2007 ,357 :2277-2284.
  • 5Brody AS, Frush DP, Huda W, et al. Radiation risk to children from computed tomography. Pediatrics ,2007,120:677-682.
  • 6Campbell J, Kalra MK, Rizzo S, et al. Scanning beyond anatomic limits of the thorax in chest CT: findings, radiation dose, and automatic tube current modulation. AJR,2005,185 : 1525-1530.
  • 7Donnelly LF, Emery KH, Brody AS,et al. Minimizing radiation dose for pediatric body application of single-detector helical CT: strategies at a large Children's Hospital. AJR, 2001, 176: 303 -306.
  • 8Nyman U, Ahl TL, Kristiansson M, et al. Patient-circumference- adapted dose regulation in body computed tomography, a practical and flexible formula. Acta Radiol, 2005,46:396-406.
  • 9Sigal-Clinquabre AB, Hennequin R, Abada HT, et al. Low- kilovohage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Rdiology,2004,231: 169-174.
  • 10Thomas CK, Muhlenbruch G, Wildberger JE, et al. Coronary artery calcium scoring with muhislice computed tomography: in vitro assessment of a low tube voltage protocol. Invest Radiol, 2006, 41:668-673.

共引文献79

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部