摘要
设计了基于三级半导体制冷的氮气J-T效应预冷器,研制了直径3 mm的冷冻探针,搭建了基于半导体制冷预冷的氮气冷冻刀降温实验系统。实验结果表明,本系统中冷冻刀在氮气初始压力8.5 MPa下节流温度可达-98.3℃,在初始压力为12 MPa下,节流后的氮气最低温度可达-120℃;冷冻刀的冷量随刀头恒定制冷温度的不同而不同,在氮气初始压力为8.5 MPa下,其制冷温度恒定为-47℃时,冷量为30 W,可满足杀死细胞的要求;水中冻结实验表明,氮气初始压力越大,冷冻刀冷量越大。
The nitrogen J-T effect pre-eooler was designed based on three-level semiconductor refrigera- tion, the cryoprobe of 3 mm diameter was developed and the nitrogen cryoprobe cooling system was set up based on the semiconductor pre-cooler. The experimental results showed that under the initial pressure of 8.5 MPa, the throttle temperature of the cryoprobe could reach to - 98.3 ℃. While under the initial pres- sure of 12 MPa, the throttle temperature of nitrogen could reach to -120 ℃. The cooling capacity of the cryoprobe was different when the constant refrigeration temperature changed. Under the pressure of 8.5 MPa, the cooling capacity was 30 W when the constant refrigeration temperature was -47 ℃ , meeting the requirements of killing cells. Freezing experiment in water also indicated that the cooling capacity increased with the increase of the initial pressure of the nitrogen.
出处
《低温工程》
CAS
CSCD
北大核心
2014年第1期50-54,共5页
Cryogenics
关键词
冷冻治疗
冷冻刀
J-T效应
预冷
半导体制冷
cryotherapy
cryoprobe
J-T effect
pre-eooling
semiconductor refrigeration