Sysmex XE-2100血液分析仪临床应用评价 被引量:12

Clinical application evaluation of XE-2100 hematology analyzer
下载PDF
导出
摘要 目的系统评价Sysmex XE-2100血液分析仪的性能。方法参照CLSI、ICSH相关文件,对XE-2100血液分析仪的精密度、携带污染率、稳定性、线性、可比性、白细胞分类相关性、正确度进行评价。结果 Sysmex XE-2100血液分析仪的精密度、携带污染率均在仪器要求范围内,标本在4℃保存条件下全血细胞计数与分类有良好的稳定性,WBC、RBC、HGB、PLT的线性范围宽,不同进样模式间的结果偏差均小于1%,与LH780、XT-1800i的检测结果具有良好的可比性,白细胞分类与人工分类具有良好的相关性,WBC、RBC、HGB、HCT、PLT正确度的偏移均小于1/4CLIA'88允许总误差。结论 XE-2100五分类血液分析仪的性能评价均达到了其技术指标,能满足临床对血液分析的需求。 Objective To evaluate the performance of Sysmex XE-2100 automated hematology analyzer. Methods According to CLSI, ICSH relevant documents, the performance of Sysmex XE-2100 was evaluated for precision, carryover rate, stability, lin-earity, comparability, correlation of leukocyte classification and accuracy. Results The results of precision and carryover rate were within the required range of instrument, complete blood count parameters and classification had good stability when the specimen were stored at 4℃. The WBC, RBC, HGB and PLT had wide linear range. The deviation of results detected with different measure modes were less than 1%. Compared with Beckman coulter LH780 and Sysmex XT-1800i, Sysmex XE-2100 had good comparabil-ity, leukocyte classification and artificial classification had good correlation. The bias of accuracy was less than a quarter CLIA'88 allow total error. Conclusion All the performance of Sysmex XE-2100 automated hematology analyzer has reached the technical indicators and meet the clinical demand for blood analysis.
出处 《实验与检验医学》 CAS 2014年第1期26-29,共4页 Experimental and Laboratory Medicine
关键词 血液分析仪 应用 评价 Automated hematology analyzer Performance Evaluation
  • 相关文献

参考文献9

  • 1彭黎明,丛玉隆.我国血细胞自动分析中存在的问题及对策[J].中华检验医学杂志,2005,28(9):885-887. 被引量:43
  • 2WS/T347-2011.血液分析仪的校准指南[S].2011.
  • 3CLSI EP15-A.Clinicaland Laboratory Standards Institute.User Demonstration of Performance for Precision and Accuracy;Ap- proved Guideline[S].CLS1,2001.
  • 4International Council for Standardization in Haematology (ICSH). Protocol for evaluation of automated hematology analyzer[S],iCSH, Clin Lab Heamotal, 1984;6:69.
  • 5CLSI EP6-A.Clinicaland Laboratory Standards Institute.Evaluation of the Linearity of Quantitative Measurement Procedures : A Statis- tical Approach ;Approved Guideline[S].CLSI, 2003.
  • 6CLSI EP9-A2.Clinicaland Laboratory Standards Institute.Method Comparison and Bias Estimation Using Patient Samples;Approved Guideline-Second Edition[S].CLS1,2002.
  • 7CLSI H20-A.Clinicaland Laboratory Standards Institute.Reference Leukocyte Differential Count (Proportional) and Evaluation of In- strumental Methods;Approved Standard[S].CLSl, 1992.
  • 8Tsuda I,Hino M ,Takubo T,et al. First basic performance evalua- tion of the XE-2100 haematology analyzer [J].J Autom Methods Manag Chem, 1999,21 (4): 127-133.
  • 9Nakul-Aquaronne D, Sudaka-Sammarcelli I, Ferrero-Vacher C,et al.Evaluation of the Sysmex XE-2100 hematology analyzer in hos- pital use[J1. J Clin Lab Anal,2003,17(4):113-123.

二级参考文献9

共引文献42

同被引文献107

引证文献12

二级引证文献33

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部