期刊文献+

原子吸收光谱法测定复方氯化钠注射液中氯化钙含量 被引量:3

Determination of Calcium Chloride in Compound Sodium Chloride Injection by Atomic Absorption Spectrometry
下载PDF
导出
摘要 目的:建立原子吸收光谱法测定复方氯化钠注射液氯化钙含量的方法。方法:采用火焰原子吸收光谱法测定复方氯化钠注射液氯化钙的含量,检测波长为422.7 nm;燃气类型为空气-乙炔,燃气比为空气体积︰乙炔体积=13.5∶2;灯电流:10 mA;狭缝宽度:0.5 nm。结果:Ca2+浓度在1~9μg/mL之间,与对应吸光度值呈良好的线性关系,回收率在98.42%~100.93%,检测限为0.008206μg/mL,与药典方法含量检测结果之间的RSD值为1.21%。结论:该方法进行复方氯化钠注射液Ca2+含量的测定专属性好,精密度高,稳定性强,重复性好,可作为药典补充方法。复方氯化钠注射液CaCl2·2H2 O质量浓度测定结果为0.0329296%~0.0340398% g/mL,符合质量标准要求。 A method was established to determine the content of calcium chloride in compound sodium chloride injection .By atom-ic absorption spectrometry , the content of calcium chloride in compound sodium chloride injection was determined under the condition as follows:detection wavelength was 422.7 nm,the flame tape was air/acetylene(13.5∶2.0),the lamp current was 10 mA,slit width was 0.5 nm.Good linear relationship of concentration and absorption of Ca 2+was within the range of 1μg/mL-9μg/mL, the recovery rate range from 98.42%to 100.93%, detectability was 0.008206μg/mL,the RSD between results obtained from official method and ones obtained from comparative method is 1.21%.This method has good specificity, stability, reproducibility, and is accurate for de-tecting the content of Ca 2+,and can be supplementary method to official one .The concentration of calcium chloride in compound sodi-um chloride injection is between 0.0329296%g/mL and 0.0340398%g/mL, which meets the required quality standards .
出处 《西华大学学报(自然科学版)》 CAS 2014年第2期93-97,共5页 Journal of Xihua University:Natural Science Edition
基金 西华大学天然药物研究与工程重点实验室项目(XZD0821-09-1)
关键词 原子吸收光谱法 复方氯化钠注射液 氯化钙 含量测定 atomic absorption spectrometry compound sodium chloride injection calcium chloride content determination
  • 相关文献

参考文献6

二级参考文献8

共引文献15

同被引文献19

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部