摘要 本文用流动注射方法研究了磷钼杂多酸与罗丹明6G的荧光淬灭反应,利用此反应测定了痕量磷,磷量的线性范围在0-80ppb,检出限0.343ppb,分析速度可达60样/小时,本法应用于铜合金、钢样和锡矿中痕量磷的测定,取得满意结果。 A flow injection analysis method for trace amount of P(Ⅴ) determination has been established, based on the fluorometric quenching effect of Rhodamine 6G due to formation of an unfluorometric ion-association complex Rhodamine 6G-molybdophosphate fluorescence was measure at 555nm ( exciation at 350nm ). The calibration graph was rectilinear from 0—80ppb of phosphorus. This method has a high sensitivity up to 0.343 ppb. The method was used to determination P(Ⅴ) in copper alloy, steel sample and manganese ore.
作者 宋功武
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 1991年第3期59-61,19,共4页 Spectroscopy and Spectral Analysis