基于MCNP使用中国人体模型对全身计数器的效率模拟研究 被引量:2

A Study of Chinese Adult Counting Efficiency Calibration Factors for a Whole Body Counter by MCNP
下载PDF
导出
摘要 为了评估放射性场所工作人员体内放射性剂量,常常使用全身计数器(WBC)探测人体的放射性,其探测效率对测量结果至关重要。研究使用蒙特卡洛方法对秦山三期的WBC的2个大面积NaI(Tl)探测器人体测量效率进行模拟,获得较好结果。研究结果表明,对能量从0.08keV^2MeV范围内误差最大为26.6%,对于身高150~180 cm的人体计数率的最大误差40%。 Whole-Body Counting was used to measure radiation dose of body in order to evaluate the internal exposure from inner -radionuclides of worker in radiation field .The efficiency of WBC detector is very impor-tant for measurement .The efficiencies of the two very large NaI (Tl) detectors (7.6 127 40.6 cm) which are used with the whole -body counter were simulated by Monte Carlo technology , good results were obtained .As a result, the relative error is 26.6%for energy from 0.08keV to 2 MeV.The discrepancy of different height a-dult person (150~180 cm) was within 40%.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2013年第12期1502-1506,共5页 Nuclear Electronics & Detection Technology
基金 秦山第三核电有限公司WBC内照射剂量测量不确定度项目资助
关键词 全身计数器 蒙特卡洛 效率 放射性 whole-body counter Monte Carlo efficiency radiation
  • 相关文献

参考文献3

  • 1~Webb, J. L. , Kramer, G.H.. An evaluation of ger-manium detectors employed for the measurement of ra- dionuclides deposited in lungs using an experimental and Monte Carlo approach[ J]. Health Phys. , 2001, 81(6) : 711 -719.
  • 2Briesmeister J. F.. MCNP - A general Monte Carlo N -particle transport code version 4C [ R ]. Los Alamos National Laboratory Report LA- 13709 -M, 2000.
  • 3E. Carinou,V. Koukoulioua,M. Budayova, et al. The calculation of a size correction factor for a whole - body counter[ J]. Nuclear Instruments and Methods in Physics Research A,2007,580( 1 ) :197 -200.

同被引文献21

引证文献2

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部